Fibrin glue for opposing wound edges in "Top Hat" penetrating keratoplasty: a laboratory study.

Bahar Irit, Kaiserman Igor, Slomovic Allan, McAllum Penny, Rootman David

СтатьяПеревод в очереди

Аннотация

PURPOSE: To evaluate the advantages of fibrin glue for opposing wound edges in Top Hat penetrating keratoplasty (PKP).

METHODS

Twenty human corneoscleral rims were mounted on an artificial anterior chamber. Eight corneas underwent traditional PKP, 6 underwent Top Hat PKP, and 6 underwent Top Hat PKP by using fibrin glue for opposing wound edges. Mechanical stability was evaluated after placement of 8 and 16 interrupted sutures. Wound bursting pressure and induced astigmatism were evaluated.

RESULTS

In the traditional PKP group, wound bursting pressure was 25.2 and 59.1 mm Hg after placement of 8 and 16 sutures, respectively. In the Top Hat PKP, leakage occurred at 57.6 and 103.8 mm Hg after placement of the 8 and 16 sutures, respectively. In the Top Hat PKP + fibrin glue group, wound leakage occurred at 144.6 mm Hg after placement of the 8 sutures and at >158 mm Hg after placement of 16 sutures. The Top Hat PKP + fibrin glue group induced astigmatism of 2.5 D, whereas the traditional PKP group and the Top Hat PKP group showed an induced astigmatism of 3.1 D each.

CONCLUSIONS

The use of fibrin glue in Top Hat PKP was found to be more mechanically stable than traditional sutures.

Для цитирования: Bahar Irit, Kaiserman Igor, Slomovic Allan, McAllum Penny, Rootman David. Fibrin glue for opposing wound edges in "Top Hat" penetrating keratoplasty: a laboratory study.. Cornea. 2007. DOI: 10.1097/ico.0b013e318151f8e8.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.