Hyperboloid-Drum Microdisk Laser Biosensors for Ultrasensitive Detection of Human IgG.

Guo Zhihe, Qin Yingchun, Chen Peizong, Hu Jinliang, Zhou Yi, Zhao Xuyang, Liu Zhiran, Fei Yiyan, Jiang Xiaoshun, Wu Xiang

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

Whispering gallery mode (WGM) microresonators have been used as optical sensors in fundamental research and practical applications. The majority of WGM sensors are passive resonators that require complex systems, thereby limiting their practicality. Active resonators enable the remote excitation and collection of WGM-modulated fluorescence spectra, without requiring complex systems, and can be used as alternatives to passive microresonators. This paper demonstrates an active microresonator, which is a microdisk laser in a hyperboloid-drum (HD) shape. The HD microdisk lasers are a combination of a rhodamine B-doped photoresist and a silica microdisk. These HD microdisk lasers can be utilized for the detection of label-free biomolecules. The biomolecule concentration can be as low as 1 ag mL-1 , whereas the theoretical detection limit of the biosensor for human IgG in phosphate buffer saline is 9 ag mL-1 (0.06 aM ). Additionally, the biosensors are able to detect biomolecules in an artificial serum, with a theoretical detection limit of 9 ag mL-1 (0.06 aM ). These results are approximately four orders of magnitude more sensitive than those for the typical active WGM biosensors. The proposed HD microdisk laser biosensors show enormous detection potential for biomarkers in protein secretions or body fluids.

Для цитирования: Guo Zhihe, Qin Yingchun, Chen Peizong, Hu Jinliang, Zhou Yi, Zhao Xuyang, Liu Zhiran, Fei Yiyan, Jiang Xiaoshun, Wu Xiang. Hyperboloid-Drum Microdisk Laser Biosensors for Ultrasensitive Detection of Human IgG.. Small (Weinheim an der Bergstrasse, Germany). 2020. DOI: 10.1002/smll.202000239.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.