Surveillance of guideline practices for duodenoscope and linear echoendoscope reprocessing in a large healthcare system.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Brandabur Jack J, Leggett James E, Wang Lian, Bartles Rebecca L, Baxter Lynda, Diaz George A, Grunkemeier Gary L, Hove Shannan, Oethinger Margret
Аннотация
BACKGROUND AND AIMS: To assess the adequacy of currently recommended duodenoscope and linear echoendoscope (DLE) automatic endoscope reprocessing (AER) and high-level disinfection (HLD), we collected daily post-reprocessing surveillance cultures of 106 DLEs in 21 Providence and Affiliate Hospitals.
METHODS: Daily qualitative surveillance of dried, post-HLD DLEs was conducted for a minimum of 30 days at each facility. Positivity rates for any microbial growth and growth of high-concern pathogens were reported. Potential effects of DLE manufacturer, age, and AER processor on culture-positivity rate were assessed.
RESULTS: Microbial growth was recovered from 201 of 4032 specimens (5%) or 189 of 2238 encounters (8.4%), including 23 specimens (.6%) or 21 encounters (.9%) for a high-concern pathogen. Wide variations in culture-positivity rate were observed across facilities. No striking difference in culture-positivity rate was seen among 8 DLE models, 3 DLE manufacturers, DLE age, manual or bedside cleanser, or automatic flushing system use. However, there was suggestive evidence that Custom Ultrasonics AER (Warminster, Pa, USA) had a lower culture-positivity rate than Medivators AER (Cantel Medical Corp., Little Falls, NJ, USA) for high-concern pathogen growth (0/1079 vs 21/2735 specimens or 0/547 vs 20/1582 encounters). Two endoscopes grew intestinal flora on several occasions despite multiple HLD. No multidrug-resistant organism was detected.
CONCLUSIONS: In this multicenter DLE surveillance study, microbial growth was recovered in 5.0% of specimens (8.4% of encounters), with most being environmental microbes. Enteric bacterial flora was recovered in .6% of specimens (.9% of encounters), despite compliance with 2014 U.S. guidelines and manufacturers' recommendations for cleaning and HLD process. The observed better performance of Custom Ultrasonics AER deserves further investigation.
METHODS: Daily qualitative surveillance of dried, post-HLD DLEs was conducted for a minimum of 30 days at each facility. Positivity rates for any microbial growth and growth of high-concern pathogens were reported. Potential effects of DLE manufacturer, age, and AER processor on culture-positivity rate were assessed.
RESULTS: Microbial growth was recovered from 201 of 4032 specimens (5%) or 189 of 2238 encounters (8.4%), including 23 specimens (.6%) or 21 encounters (.9%) for a high-concern pathogen. Wide variations in culture-positivity rate were observed across facilities. No striking difference in culture-positivity rate was seen among 8 DLE models, 3 DLE manufacturers, DLE age, manual or bedside cleanser, or automatic flushing system use. However, there was suggestive evidence that Custom Ultrasonics AER (Warminster, Pa, USA) had a lower culture-positivity rate than Medivators AER (Cantel Medical Corp., Little Falls, NJ, USA) for high-concern pathogen growth (0/1079 vs 21/2735 specimens or 0/547 vs 20/1582 encounters). Two endoscopes grew intestinal flora on several occasions despite multiple HLD. No multidrug-resistant organism was detected.
CONCLUSIONS: In this multicenter DLE surveillance study, microbial growth was recovered in 5.0% of specimens (8.4% of encounters), with most being environmental microbes. Enteric bacterial flora was recovered in .6% of specimens (.9% of encounters), despite compliance with 2014 U.S. guidelines and manufacturers' recommendations for cleaning and HLD process. The observed better performance of Custom Ultrasonics AER deserves further investigation.