MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.brachy.2025.04.005

Медицинская библиотека · Brachytherapy · 2025

Clinical implementation of an open-source Monte Carlo system across multiple centers for permanent implant brachytherapy dose evaluation.

Moghadam Narjes, Akbari Fatemeh, Zhang Claire, Ouellet Samuel, McKeown Dakota, Hoseini-Ghahfarokhi Mojtaba, Parent Sandra, Carrier Jean-François, Lavallée Marie-Claude, Cygler Joanna E, Hilts Michelle, Vigneault Eric, Crook Juanita, Beaulieu Luc, Thomson Rowan M

СтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

PURPOSE: This study aims to clinically implement eb_gui, a user-friendly toolkit for Monte Carlo simulations utilizing egs_brachy, in the context of low-dose-rate (LDR) brachytherapy for prostate and breast cancers.

METHODS

A set of test cases, ranging from simple to complex scenarios including single- and multi-seed patient models, was developed for LDR brachytherapy of both breast (103Pd) and prostate (125I). Utilizing Digital Imaging and Communications in Medicine (DICOM) files, the open-source interface eb_gui was employed to compute doses. The commissioning process involved comparing eb_gui results against clinical TG-43 treatment planning system (TG43-TPS) calculations, encompassing point-by-point differences across 3D dose distributions, dose volume histograms, and dose metrics. Additionally, patient-specific dose distributions were computed using eb_gui's full-tissue models (TG186-MC) and compared against TG-43 Monte Carlo calculations (TG43-MC) across multiple cancer centers.

RESULTS

Excellent agreement was observed between TG43-TPS and TG43-MC calculated doses, with point-to-point differences of less than 1 Gy (∼1% of prescription dose) for breast and prostate cases. Comparisons between multicenter TG186-MC and TG43-MC doses highlighted discrepancies that underscore the limitations of the TG-43 formalism and affirming the necessity for a model-based dose calculation algorithm (MBDCA).

CONCLUSION

This study successfully developed a series of test cases and a commissioning workflow for implementing eb_gui in LDR brachytherapy across multiple centers. The findings underscore the potential of TG-186 MBDCA to enhance the precision of patient dosimetry and improve the accuracy of treatment outcome predictions in LDR brachytherapy. This work represents a significant step toward broader adoption of advanced dose calculation methodologies in clinical practice.

Для цитирования: Moghadam Narjes, Akbari Fatemeh, Zhang Claire, Ouellet Samuel, McKeown Dakota, Hoseini-Ghahfarokhi Mojtaba, Parent Sandra, Carrier Jean-François, Lavallée Marie-Claude, Cygler Joanna E, Hilts Michelle, Vigneault Eric, Crook Juanita, Beaulieu Luc, Thomson Rowan M. Clinical implementation of an open-source Monte Carlo system across multiple centers for permanent implant brachytherapy dose evaluation.. Brachytherapy. 2025. DOI: 10.1016/j.brachy.2025.04.005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.