DOI 10.1097/01.mat.0000196522.29376.96

Медицинская библиотека · ASAIO journal (American Society for Artificial Internal Organs : 1992) · 2006

Predicted hemodynamic benefits of counterpulsation therapy using a superficial surgical approach.

Giridharan Guruprasad A, Pantalos George M, Litwak Kenneth N, Spence Paul A, Koenig Steven C

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

A volume-displacement counterpulsation device (CPD) intended for chronic implantation via a superficial surgical approach is proposed. The CPD is a pneumatically driven sac that fills during native heart systole and empties during diastole through a single, valveless cannula anastomosed to the subclavian artery. Computer simulation was performed to predict and compare the physiological responses of the CPD to the intraaortic balloon pump (IABP) in a clinically relevant model of early stage heart failure. The effect of device stroke volume (0-50 ml) and control modes (timing, duration, morphology) on landmark hemodynamic parameters and the LV pressure-volume relationship were investigated. Simulation results predicted that the CPD would provide hemodynamic benefits comparable to an IABP as evidenced by up to 25% augmentation of peak diastolic aortic pressure, which increases diastolic coronary perfusion by up to 34%. The CPD may also provide up to 34% reduction in LV end-diastolic pressure and 12% reduction in peak systolic aortic pressure, lowering LV workload by up to 26% and increasing cardiac output by up to 10%. This study demonstrated that the superficial CPD technique may be used acutely to achieve similar improvements in hemodynamic function as the IABP in early stage heart failure patients.

Для цитирования: Giridharan Guruprasad A, Pantalos George M, Litwak Kenneth N, Spence Paul A, Koenig Steven C. Predicted hemodynamic benefits of counterpulsation therapy using a superficial surgical approach.. ASAIO journal (American Society for Artificial Internal Organs : 1992). 2006. DOI: 10.1097/01.mat.0000196522.29376.96.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.