Flow modulation comprehensive two-dimensional gas chromatography-mass spectrometry using ≈4 mL min(-1) gas flows.

Franchina Flavio A, Maimone Mariarosa, Tranchida Peter Q, Mondello Luigi

СтатьяПеревод в очереди

Аннотация

The main objective of the herein described research was focused on performing satisfactory flow modulation (FM), in comprehensive two-dimensional gas chromatography-mass spectrometry (GC×GC-MS), using an MS-compatible second-dimension gas flow of approx. 4 mL min(-1). The FM model used was based on that initially proposed by Seeley et al. [3]. The use of limited gas flows was enabled through fine tuning of the FM parameters, in particular the duration of the re-injection (or flushing) process. Specifically, the application of a long re-injection period (i.e., 700 ms) enabled efficient accumulation-loop flushing with gas flows of about 4 mL min(-1). It was possible to apply such extended re-injection periods by using different restrictor lengths in the connections linking the modulator to the auxiliary pressure source. FM GC×GC-MS applications were performed on a mixture containing C9-10 alkanes, and on a sample of essential oil. GC×GC-MS sensitivity was compared with that attained by using conventional GC-MS analysis, in essential oil applications. It was observed that signal intensities were, in general, considerably higher in the FM GC×GC-MS experiments.

Для цитирования: Franchina Flavio A, Maimone Mariarosa, Tranchida Peter Q, Mondello Luigi. Flow modulation comprehensive two-dimensional gas chromatography-mass spectrometry using ≈4 mL min(-1) gas flows.. Journal of chromatography. A. 2016. DOI: 10.1016/j.chroma.2016.02.041.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.