Mesh-covered (Roadsaver) stent as a new treatment modality for symptomatic or high-risk carotid stenosis.

Machnik Roman, Paluszek Piotr, Tekieli Łukasz, Dzierwa Karolina, Maciejewski Damian, Trystuła Mariusz, Brzychczy Andrzej, Banaszkiewicz Krzysztof, Musiał Robert, Pieniążek Piotr

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Prevention of periprocedural stroke has a crucial role in carotid artery stenting (CAS) procedures.

AIM: To assess retrospectively 30-day safety and effectiveness of 41 procedures of internal and common carotid artery stenting using the Roadsaver double nitinol layer micromesh stent in 40 non-consecutive patients with symptomatic or high-risk carotid artery stenosis.

MATERIAL AND

METHODS

The patients were men (n = 31) and women (n = 9); mean age was 67.8 ±7.9 years. Femoral access was used in 39 cases, whereas radial access was used in 2. Proximal (n = 27) or distal (n = 14) embolic neuroprotection was used.

RESULTS

The Roadsaver stents (nominal diameter 7, 8 or 9 mm, length 25 or 30 mm) were implanted successfully in all cases. One minor stroke occurred after common carotid artery intubation with a guiding catheter (before stent deployment) and one transient postprocedural ischemic attack (TIA) of the ipsilateral cerebral hemisphere was observed. Internal/common carotid artery stenosis severity was evaluated by duplex Doppler. Maximal peak systolic velocity (PSV) before CAS was in the range: 2.0-7.0 m/s, mean: 3.9 ±1.0 m/s, at 24-48 h after stenting mean PSV was 1.1 ±0.4 m/s (p < 0.05), and at 30 days 1.1 ±0.3 m/s (p < 0.05). Maximal end-diastolic velocity (EDV) was 0.85-3.5 m/s, mean 1.4 ±0.5 m/s, at 24-48 h after stenting mean EDV was 0.3 ±0.1 m/s (p < 0.05), and at 30 days 0.4 ±0.1 m/s (p < 0.05). No restenosis or thrombosis was observed. Angiographic stenosis decreased from 82.9 ±9.1% (range: 61-97%) to 19.3 ±7.3% (range: 0-34%) (p < 0.05).

CONCLUSIONS

The CAS using the Roadsaver stent seems to be safe and effective. Further studies involving larger patient populations and longer follow-up are needed.

Для цитирования: Machnik Roman, Paluszek Piotr, Tekieli Łukasz, Dzierwa Karolina, Maciejewski Damian, Trystuła Mariusz, Brzychczy Andrzej, Banaszkiewicz Krzysztof, Musiał Robert, Pieniążek Piotr. Mesh-covered (Roadsaver) stent as a new treatment modality for symptomatic or high-risk carotid stenosis.. Postepy w kardiologii interwencyjnej = Advances in interventional cardiology. 2017. DOI: 10.5114/pwki.2017.68139.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.