Artificial intelligence assists operators in real-time detection of focal liver lesions during ultrasound: A randomized controlled study.

Tiyarattanachai Thodsawit, Apiparakoon Terapap, Chaichuen Oracha, Sukcharoen Sasima, Yimsawad Sirinda, Jangsirikul Sureeporn, Chaikajornwat Jukkaphop, Siriwong Nanicha, Burana Chuti, Siritaweechai Natakorn, Atipas Kawin, Assawamasbunlue Nongnapas, Tovichayathamrong Punyot, Obcheuythed Pitchanun, Somvanapanich Pochara, Geratikornsupuk Nopavut, Anukulkarnkusol Nopporn, Sarakul Pamornmas, Tanpowpong Natthaporn, Pinjaroen Nutcha, Kerr Stephen J, Rerknimitr Rungsun, Marukatat Sanparith, Chaiteerakij Roongruedee

Рандомизированное исследованиеСтатьяПеревод в очереди

Аннотация

PURPOSE: Detection of hepatocellular carcinoma (HCC) is crucial during surveillance by ultrasound. We previously developed an artificial intelligence (AI) system based on convolutional neural network for detection of focal liver lesions (FLLs) in ultrasound. The primary aim of this study was to evaluate whether the AI system can assist non-expert operators to detect FLLs in real-time, during ultrasound examinations.

METHOD: This single-center prospective randomized controlled study evaluated the AI system in assisting non-expert and expert operators. Patients with and without FLLs were enrolled and had ultrasound performed twice, with and without AI assistance. McNemar's test was used to compare paired FLL detection rates and false positives between groups with and without AI assistance.

RESULTS

260 patients with 271 FLLs and 244 patients with 240 FLLs were enrolled into the groups of non-expert and expert operators, respectively. In non-experts, FLL detection rate in the AI assistance group was significantly higher than the no AI assistance group (36.9 % vs 21.4 %, p < 0.001). In experts, FLL detection rates were not significantly different between the groups with and without AI assistance (66.7 % vs 63.3 %, p = 0.32). False positive detection rates in the groups with and without AI assistance were not significantly different in both non-experts (14.2 % vs 9.2 %, p = 0.08) and experts (8.6 % vs 9.0 %, p = 0.85).

CONCLUSIONS

The AI system resulted in significant increase in detection of FLLs during ultrasound examinations by non-experts. Our findings may support future use of the AI system in resource-limited settings where ultrasound examinations are performed by non-experts. The study protocol was registered under the Thai Clinical Trial Registry (TCTR20201230003), which is part of the WHO ICTRP Registry Network. The registry can be accessed via the following URL: https://trialsearch.who.int/Trial2.aspx?TrialID=TCTR20201230003.

Для цитирования: Tiyarattanachai Thodsawit, Apiparakoon Terapap, Chaichuen Oracha, Sukcharoen Sasima, Yimsawad Sirinda, Jangsirikul Sureeporn, Chaikajornwat Jukkaphop, Siriwong Nanicha, Burana Chuti, Siritaweechai Natakorn, Atipas Kawin, Assawamasbunlue Nongnapas, Tovichayathamrong Punyot, Obcheuythed Pitchanun, Somvanapanich Pochara, Geratikornsupuk Nopavut, Anukulkarnkusol Nopporn, Sarakul Pamornmas, Tanpowpong Natthaporn, Pinjaroen Nutcha, Kerr Stephen J, Rerknimitr Rungsun, Marukatat Sanparith, Chaiteerakij Roongruedee. Artificial intelligence assists operators in real-time detection of focal liver lesions during ultrasound: A randomized controlled study.. European journal of radiology. 2023. DOI: 10.1016/j.ejrad.2023.110932.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.