Acute oxygenator failure: a new presentation of heparin-induced thrombocytopenia in a patient undergoing venovenous extracorporeal membrane oxygenation support.

Ratzlaff Robert A, Ripoll Juan G, Kassab Lena L, Diaz-Gomez Jose L

Клинический случайСтатьяПеревод в очереди

Аннотация

A 58-year-old man with medical history of thrombocytopenia was admitted to an outside hospital for a 6-day history of worsening dyspnoea requiring mechanical ventilator support. He was transferred to our institution for extracorporeal membrane oxygenation (ECMO) given his refractory hypoxaemia. On arrival, H1N1 influenza virus was confirmed and all measures to improve oxygenation were ineffective. Thus, the decision was made to start venovenous (VV)-ECMO. Although a low baseline platelet count was recognised (60-70×109/L), a sudden further decrease occurred (30×109/L) and platelet transfusion was initiated. A substantial increase in the pressure across the ECMO oxygenator was identified, and the diagnosis of type II heparin-induced thrombocytopenia was suspected and confirmed. Heparin was discontinued, the oxygenator was exchanged and argatroban was used for anticoagulation. After 28 days on VV-ECMO support, the decision was made to withdraw organ support in conjunction with the patient and family wishes.

Для цитирования: Ratzlaff Robert A, Ripoll Juan G, Kassab Lena L, Diaz-Gomez Jose L. Acute oxygenator failure: a new presentation of heparin-induced thrombocytopenia in a patient undergoing venovenous extracorporeal membrane oxygenation support.. BMJ case reports. 2016. DOI: 10.1136/bcr-2016-218179.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.