Jones Ryan M, Huang Yuexi, Meng Ying, Scantlebury Nadia, Schwartz Michael L, Lipsman Nir, Hynynen Kullervo
Аннотация
BACKGROUND: Transcranial magnetic resonance-guided focused ultrasound (TcMRgFUS) systems currently employ computed tomography (CT)-based aberration corrections, which may provide suboptimal trans-skull focusing.
OBJECTIVES: The objective of this study was to evaluate a contrast agent microbubble imaging-based transcranial focusing method, echo-focusing (EF), during TcMRgFUS for essential tremor.
METHODS: A clinical trial of TcMRgFUS thalamotomy using EF for the treatment of essential tremor was conducted (NCT03935581; funded by InSightec [Tirat Carmel, Israel]). Patients (n = 12) were injected with Definity (Lantheus Medical Imaging, North Billerica, MA) microbubbles, and EF was performed using a research feature add-on to a commercial TcMRgFUS system (ExAblate Neuro, InSightec). Subablative thermal sonications carried out using (1) EF and (2) CT-based aberration corrections were compared via magnetic resonance thermometry, and the optimal focusing method for each patient was employed for TcMRgFUS thalamotomy.
RESULTS: EF aberration corrections provided increased sonication efficiency, decreased focal size, and equivalent targeting accuracy relative to CT-based focusing. EF aberration corrections were employed successfully for lesion formation in all 12 patients, 3 of whom had previously undergone unsuccessful TcMRgFUS thalamotomy via CT-based focusing. There were no adverse events related directly to the EF procedure.
CONCLUSIONS: EF is feasible and appears safe during TcMRgFUS thalamotomy for essential tremor and improves on the trans-skull focal quality provided by existing CT-based focusing methods. © 2020 International Parkinson and Movement Disorder Society.
OBJECTIVES: The objective of this study was to evaluate a contrast agent microbubble imaging-based transcranial focusing method, echo-focusing (EF), during TcMRgFUS for essential tremor.
METHODS: A clinical trial of TcMRgFUS thalamotomy using EF for the treatment of essential tremor was conducted (NCT03935581; funded by InSightec [Tirat Carmel, Israel]). Patients (n = 12) were injected with Definity (Lantheus Medical Imaging, North Billerica, MA) microbubbles, and EF was performed using a research feature add-on to a commercial TcMRgFUS system (ExAblate Neuro, InSightec). Subablative thermal sonications carried out using (1) EF and (2) CT-based aberration corrections were compared via magnetic resonance thermometry, and the optimal focusing method for each patient was employed for TcMRgFUS thalamotomy.
RESULTS: EF aberration corrections provided increased sonication efficiency, decreased focal size, and equivalent targeting accuracy relative to CT-based focusing. EF aberration corrections were employed successfully for lesion formation in all 12 patients, 3 of whom had previously undergone unsuccessful TcMRgFUS thalamotomy via CT-based focusing. There were no adverse events related directly to the EF procedure.
CONCLUSIONS: EF is feasible and appears safe during TcMRgFUS thalamotomy for essential tremor and improves on the trans-skull focal quality provided by existing CT-based focusing methods. © 2020 International Parkinson and Movement Disorder Society.