MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jtcvs.2007.03.003

Медицинская библиотека · The Journal of thoracic and cardiovascular surgery · 2007

A reinforced sternal wiring technique for transverse thoracosternotomy closure in bilateral lung transplantation: from biomechanical test to clinical application.

Oto Takahiro, Venkatachalam Ramkumar, Morsi Yos S, Marasco Silvana, Pick Adrian, Rabinov Marc, Rosenfeldt Franklin

Оценочное исследованиеСтатьяПеревод в очереди

Аннотация

OBJECTIVES

A high incidence of failure of transverse thoracosternotomy closure, involving the loops of wire cutting through the sternum, remains a significant morbidity after bilateral lung transplantation. We postulated that placing peristernal wires inside the usual longitudinal wires could prevent the longitudinal wires from cutting through the sternum. The aims of this study were to investigate the biomechanical and clinical efficacy of the proposed reinforced sternal closure technique.

METHODS

In vitro, 24 artificial sternal models were wired with the reinforced or conventional wiring techniques and were tested either by means of longitudinal distraction or anterior-posterior shear (n = 6 per group). In vivo, the 6-month outcomes of 70 bilateral lung transplantations, including 27 reinforced and 43 conventional wiring techniques, were assessed.

RESULTS

Reinforced wiring was stronger than conventional wiring for both longitudinal distraction (yield load: 585 +/- 60 vs 334 +/- 21 N [P = .03]; maximum load: 807 +/- 60 vs 525 +/- 34 N [P = .03]; postyield stiffness: 91.0 +/- 22.0 vs 32.8 +/- 11.8 N/mm [P = .04]) and anterior-posterior shear (yield load: 405 +/- 9 vs 364 +/- 16 N [P = .03]; postyield stiffness: 47.4 +/- 6.1 vs 27.5 +/- 5.1 N/mm [P = .04]). In multivariate analysis, the use of the conventional wiring technique (odds ratio, 5.38; P = .04) and osteoporosis (odds ratio, 18.31; P = .0005) were significant risk factors associated with sternal dehiscence. In the patients with osteoporosis (n = 25), the incidence of sternal dehiscence in the reinforced wiring group (4/16 [25%]) was significantly lower than that in the conventional wiring group (7/9 [78%], P = .02).

CONCLUSION

Osteoporosis is a significant risk factor for sternal dehiscence after bilateral lung transplantation. The new reinforced sternal wiring technique provides biomechanically superior fixation of the sternum and clinically reduces the incidence of sternal dehiscence in high-risk osteoporotic patients undergoing bilateral lung transplantation.

Для цитирования: Oto Takahiro, Venkatachalam Ramkumar, Morsi Yos S, Marasco Silvana, Pick Adrian, Rabinov Marc, Rosenfeldt Franklin. A reinforced sternal wiring technique for transverse thoracosternotomy closure in bilateral lung transplantation: from biomechanical test to clinical application.. The Journal of thoracic and cardiovascular surgery. 2007. DOI: 10.1016/j.jtcvs.2007.03.003.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.