Li Tian-di, Zhou Wei, Yi Juan, Zhang Wen, Lin Yi-Ran, Li Shuang-feng
Аннотация
OBJECTIVE: To establish a gas chromatographic method for determination of halogenated alkanes and aromatic hydrocarbons including trichloromethane, 1,2-dichloroethane, carbon tetrachloride, benzene, toluene, ethylbenzene and xylene in the air of workplace.
METHODS: After the air samples collected with activated carbon tubes and desorbed with CS(2), the target toxicants were separated with FFAP capillary columns and detected with flame ionization detector.
RESULTS: The coefficient of correlation was above 0.999 and the lowest detectable concentrations were 0.2 ∼ 3.6 mg/m(3) with the RSD of 1.2% ∼ 4.6%. The desorption efficiencies was 94.9% ∼ 100.7%.
CONCLUSION: The method shows lower detection limit, high accuracy and precision. It is feasible for determination of the seven halogenated alkanes and aromatic hydrocarbons in the air of workplace.
METHODS: After the air samples collected with activated carbon tubes and desorbed with CS(2), the target toxicants were separated with FFAP capillary columns and detected with flame ionization detector.
RESULTS: The coefficient of correlation was above 0.999 and the lowest detectable concentrations were 0.2 ∼ 3.6 mg/m(3) with the RSD of 1.2% ∼ 4.6%. The desorption efficiencies was 94.9% ∼ 100.7%.
CONCLUSION: The method shows lower detection limit, high accuracy and precision. It is feasible for determination of the seven halogenated alkanes and aromatic hydrocarbons in the air of workplace.