Novel electrode design for potentially painless internal defibrillation also allows for successful external defibrillation.

Jayanti Venku, Zviman Menekhem M, Nazarian Saman, Halperin Henry R, Berger Ronald D

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Implantable cardioverter defibrillators (ICDs) save lives, but the defibrillation shocks delivered by these devices produce substantial pain, presumably due to skeletal muscle activation. In this study, we tested an electrode system composed of epicardial panels designed to shield skeletal muscles from internal defibrillation, but allow penetration of an external electric field to enable external defibrillation when required.

METHODS AND

RESULTS

Eleven adult mongrel dogs were studied under general anesthesia. Internal defibrillation threshold (DFT) and shock-induced skeletal muscle force at various biphasic shock strengths were compared between two electrode configurations: (1) a transvenous coil placed in the right ventricle (RV) as cathode and a dummy can placed subcutaneously in the left infraclavicular fossa as anode (control configuration) and (2) RV coil as cathode and the multielectrode epicardial sock with the panels connected together as anode (sock-connected). External DFT was also tested with these electrode configurations, as well as with the epicardial sock present, but with panels disconnected from each other (sock-disconnected). Internal DFT was higher with sock-connected than control (24 +/- 7 J vs. 16 +/- 6 J, P < 0.02), but muscle contraction force at DFT was greatly reduced (1.3 +/- 1.3 kg vs. 10.6 +/- 2.2 kg, P < 0.0001). External defibrillation was never successful, even at 360 J, with sock-connected, while always possible with sock-disconnected.

CONCLUSION

Internal defibrillation with greatly reduced skeletal muscle stimulation can be achieved using a novel electrode system that also preserves the ability to externally defibrillate when required. This system may provide a means for painless ICD therapy.

Для цитирования: Jayanti Venku, Zviman Menekhem M, Nazarian Saman, Halperin Henry R, Berger Ronald D. Novel electrode design for potentially painless internal defibrillation also allows for successful external defibrillation.. Journal of cardiovascular electrophysiology. 2007. DOI: 10.1111/j.1540-8167.2007.00936.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.