Objective assessment of the combined effect of exomass-related- and motion artefacts in cone beam CT.

Oliveira Matheus Lima, Candemil Amanda P, Freitas Deborah Q, Haiter-Neto Francisco, Wenzel Ann, Spin-Neto Rubens

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

To assess quantitatively the combined effect of exomass-related- and motion artefacts on voxel value parameters in cone beam CT (CBCT).

METHODS

A cylindrical phantom was manufactured, containing 21 tubes filled with a radiopaque solution, allowing the inclusion of three titanium implants in the periphery to induce exomass-related artefacts. The phantom was mounted on a robot simulating 0.75-, 1.5-, and 3 mm movements (nodding/lateral rotation/tremor). CBCT images with/without exomass and with/without movements were acquired in duplicate in three units: Cranex 3Dx, Orthophos SL-3D, and X1 (with motion-artefact correction). A cylindrical volume of interest was defined in each tube and voxel value mean and standard deviation were assessed. For each CBCT volume, the 21 mean voxel values were averaged providing the overall mean voxel value (MVV), and the standard deviation (among the 21 values) was calculated providing overall voxel value inhomogeneity (VVI). The standard deviation from each of the 21 volumes-of-interest were averaged, providing overall image noise (IN). The effect of the diverse tested situations was inferred from a repeated-measures analysis of variance, followed by Sidak's test (α = 0.05).

RESULTS

Overall, images acquired with exomass showed significantly (p ≤ 0.05) lower MVV, and higher VVI and IN. Motion artefacts aggravated exomass-related alterations. MVV and VVI were mostly affected by 3 mm nodding movements. Motion-artefact correction eliminated the deleterious effect of movement.

CONCLUSIONS

CBCT voxel-value parameters are altered by exomass-related artefacts, and this finding is aggravated in the presence of motion artefacts. Motion-artefact correction effectively eliminated the deleterious impact of movement.

Для цитирования: Oliveira Matheus Lima, Candemil Amanda P, Freitas Deborah Q, Haiter-Neto Francisco, Wenzel Ann, Spin-Neto Rubens. Objective assessment of the combined effect of exomass-related- and motion artefacts in cone beam CT.. Dento maxillo facial radiology. 2020. DOI: 10.1259/dmfr.20200255.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.