DOI 10.1002/jssc.200800758

Медицинская библиотека · Journal of separation science · 2009

Development and validation of an HPLC method for the determination of seven penicillin antibiotics in veterinary drugs and bovine blood plasma.

Samanidou Victoria F, Giannakis Dimitrios E, Papadaki Antonia

СтатьяВалидационное исследованиеПеревод в очереди

Аннотация

Herein a quantitative method for the determination of seven penicillins in bovine plasma and veterinary drugs has been developed. Amoxicillin (AMO), ampicillin (AMP), penicillin G (PENG), penicillin V (PENV), oxacillin (OXA), cloxacillin (CLO) and dicloxacillin (DICLO) were separated on a Perfectsil ODS-2 (250 x 4 mm, 5 microm) column, using gradient elution, with a mobile phase of 0.1% v/v TFA and ACN-methanol (90:10 v/v). PDA detection was used at 240 nm. Penicillins were isolated from bovine plasma by SPE on Lichrolut RP-18 cartridges with mean recoveries from 85.7 to 113.5%. Colchicine (3 ng/microL) was used as an internal standard. The developed method was validated in terms of selectivity, linearity, accuracy, precision, stability and sensitivity. Repeatability (n = 5) and between-day precision (n = 5) revealed RSD < 12%. The detection limits in the bovine plasma were estimated as 18 ng for AMO and AMP, 25 for PENG, PENV and OXA, 3 ng for CLO and 12 ng for DICLO. Spiked plasma samples were stable for 1 wk, except for AMP and CLO, which were stable for 3 wk and OXA for 4 wk. AMO, PENG and PENV were stable for two freeze-thaw cycles, OXA, CLO and DICLO for four, while AMP only for one.

Для цитирования: Samanidou Victoria F, Giannakis Dimitrios E, Papadaki Antonia. Development and validation of an HPLC method for the determination of seven penicillin antibiotics in veterinary drugs and bovine blood plasma.. Journal of separation science. 2009. DOI: 10.1002/jssc.200800758.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.