DOI 10.1038/s41391-025-00984-7
Feasibility and safety study of the Flostent™ system (RAPID-I).
Аннотация
INTRODUCTION
This study aimed to evaluate the feasibility and safety of the FloStent™, a novel nitinol prostatic stent, in men with LUTS secondary to BPH.
MATERIALS AND
METHODS
This multi-center, single-arm, one-year study included men aged ≥45 with BPH. Participants underwent FloStent™ implantation using a flexible cystoscope. The primary endpoints were device implantability, tolerability, and device retrievability. Secondary assessments included International Prostate Symptom Score (IPSS), peak urinary flow rate (Qmax), post-void residual (PVR), and adverse events (AEs) at follow-ups over 52 weeks.
RESULTS
The study enrolled 15 participants with a mean age of 58.1 years. All attempts at implantation were successful. Postoperative catheterization was not required in any patient. Significant improvements were observed in IPSS and Qmax. Thirteen AEs were reported, all resolved, with no serious complications. Device retrieval was safely performed using standard urological equipment up to one-year post-implantation.
CONCLUSIONS
This preliminary study indicates that the FloStent™ is a safe and effective treatment option for BPH, providing significant symptom relief and improved urinary function with a favorable risk profile. Further research, including larger, randomized controlled trials, are needed to validate these findings and establish the long-term efficacy and safety of the FloStent™ device.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.