MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/aor.14179

Медицинская библиотека · Artificial organs · 2022

Hemodynamics of small arterial return cannulae for venoarterial extracorporeal membrane oxygenation.

Stephens Andrew F, Wickramarachchi Avishka, Burrell Aidan J C, Bellomo Rinaldo, Raman Jaishankar, Gregory Shaun D

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Venoarterial extracorporeal membrane oxygenation (ECMO) provides mechanical support for critically ill patients with cardiogenic shock. Typically, the size of the arterial return cannula is chosen to maximize flow. However, smaller arterial cannulae may reduce cannula-related complications and be easier to insert. This in vitro study quantified the hemodynamic effect of different arterial return cannula sizes in a simulated acute heart failure patient.

METHODS

Baseline support levels were simulated with a 17 Fr arterial cannula in an ECMO circuit attached to a cardiovascular simulator with targeted partial (2.0 L/min ECMO flow, 60-65 mm Hg mean aortic pressure-MAP) and targeted full ECMO support (3.5 L/min ECMO flow and 70-75 mm Hg MAP). Return cannula size was varied (13-21 Fr), and hemodynamics were recorded while keeping ECMO pump speed constant and adjusting pump speed to restore desired support levels.

RESULTS

Minimal differences in hemodynamics were found between cannula sizes in partial support mode. A maximum pump speed change of +600 rpm was required to reach the support target, and arterial cannula inlet pressure varied from 79 (21 Fr) to 224 mm Hg (13 Fr). The 15 Fr arterial cannula could provide the target full ECMO support at the targeted hemodynamics; however, the 13 Fr cannula could not due to the high resistance associated with the small diameter.

CONCLUSIONS

A 15 Fr arterial return cannula provided targeted partial and full ECMO support to a simulated acute heart failure patient. Balancing reduced cannula size and ECMO support level may improve patient outcomes by reducing cannula-related adverse events.

Для цитирования: Stephens Andrew F, Wickramarachchi Avishka, Burrell Aidan J C, Bellomo Rinaldo, Raman Jaishankar, Gregory Shaun D. Hemodynamics of small arterial return cannulae for venoarterial extracorporeal membrane oxygenation.. Artificial organs. 2022. DOI: 10.1111/aor.14179.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.