Percutaneous ablation of colorectal liver metastases: a comparison between the outcomes of grayscale US guidance and Sonazoid CEUS Kupffer phase guidance using propensity score matching.
Аннотация
PURPOSE: To assess the utility of Sonazoid contrast-enhanced ultrasound (CEUS) for guiding percutaneous microwave ablation (MWA) for colorectal liver metastases (CRLMs).
MATERIALS AND
METHODS
The medical records of patients who had undergone ultrasound (US)-guided percutaneous MWA between July 2020 and June 2022, were reviewed. Propensity score matching (PSM) with a ratio of 1:1 was used to balance the potential bias between the grayscale US-guided and Sonazoid CEUS-guided groups. Local tumor progression (LTP), intrahepatic recurrence (IR), and complication rates were compared between the two groups.
RESULTS
Of 252 patients enrolled, 247 achieved complete ablation, and the technical effectiveness was 98.0% (247/252). Of these 247 patients, 158 were in the grayscale US-guided group and 89 in the Sonazoid CEUS-guided group. The median follow-up period was 14.6 months. After PSM, there were no significant differences in LTP, IR, or complication rates between the two groups (p = 0.100, p = 0.511, p > 0.99, respectively). Multivariate analysis identified tumor size ≥ 3 cm (hazard ratio [HR], 7.945; 95% CI, 2.591-24.370; p < 0.001), perivascular (HR, 2.331; 95% CI, 1.068-5.087; p = 0.034), and tumor depth > 8 cm (HR, 3.194; 95% CI, 1.439-7.091; p = 0.004) as significant factors associated with LTP. For tumors with poor vision on grayscale US, Sonazoid CEUS-guided ablation achieved a better LTP rate than grayscale US-guided ablation (3.7% vs.14.8%, p = 0.032).
CONCLUSION
For tumors with poor vision on grayscale US, Sonazoid CEUS guidance is recommended for better local tumor control.