A constricting band: an unusual cause of incomplete expansion of Amplatzer septal occluder device.

Trehan Vijay, Mehta Vimal, Mukhopadhyay Saibal, Yusuf Jamal, Arora Ramesh, Kaul U A

СтатьяПеревод в очереди

Аннотация

Various problems and complications have been reported during transcatheter closure of ostium secundum atrial septal defect with Amplatzer septal occluder. We report an unusual problem that was responsible for incomplete expansion of the waist of the device, not yet reported in world literature. A polyester band in continuation with the polyester mesh was found constricting the waist of the device, which after being cut led to successful redeployment of the device.

Для цитирования: Trehan Vijay, Mehta Vimal, Mukhopadhyay Saibal, Yusuf Jamal, Arora Ramesh, Kaul U A. A constricting band: an unusual cause of incomplete expansion of Amplatzer septal occluder device.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2004. DOI: 10.1002/ccd.10771.