DOI 10.1097/mat.0000000000000479
Influence of IABP-Induced Abdominal Occlusions on Aortic Hemodynamics: A Patient-Specific Computational Evaluation.
Аннотация
Intraaortic balloon pump (IABP) is used as temporary mechanical assistance in case of cardiovascular diseases, even if different hemodynamic problems and, thus, clinical complications may happen, such as the decrease of visceral perfusion. A computational fluid dynamic (CFD) study was carried out to investigate the effects of different IABP-induced abdominal occlusions on patient-specific aortic flow. Two possible sizes (25 and 34 cm) and two locations (2 and 3 cm) of the balloon were compared, modeling four abdominal occlusions and numerically reproducing IAB inflation/deflation behavior. The results highlighted that the perfusion in renal, mesenteric, and iliac arteries decreases when the abdominal occlusion increases with balloon inflation. The study illustrates also how the balloon size affects the flow in aorta vessels in both locations, and that the positioning is of little relevance for the 34 cm balloon, whereas it influences the aortic flow very much in case of 25 cm IAB. This analysis demonstrates how the IAB-induced occlusion may vary the abdominal circulation; therefore, the correct size and positioning are emphasized for patient's outcome.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.