Аннотация

INTRODUCTION

Mechanical circulatory support (MCS) devices are increasingly used as a treatment option in resuscitation or in patients with cardiogenic shock (CS). Prophylactic implantation in high-risk percutaneous coronary interventions (HRPCI) is another upcoming indication. The i-cor ECG-synchronized cardiac assist device combines the hemodynamic support of a veno-arterial extracorporeal membrane oxygenation (VA-ECMO) with the ability to generate a pulsatile flow and thus decreasing adverse effects of VA-ECMO on myocardial function. Aim of this study was to obtain data concerning feasibility, safety and outcomes in both indications.

METHODS

A total of 47 patients (34 HRPCI, 13 CS) were included in nine German centers and participated in this study. Demographic and clinical parameters, procedural as well as follow-up data were prospectively recorded and analyzed.

RESULTS

Device implantation and initiation of ECG-synchronized cardiac assist was technical successful in all cases and no failures of the consoles or disposable parts were observed. Furthermore, intended percutaneous coronary interventions and successful weaning from cardiac assist was achieved in 97.1% of HRPCI patients. We observed a 30d-survival of 94.1% in the HRPCI group and 69.2% in the CS group. Main complications in both groups were bleeding events (14.7% HRPCI, 23.1% CS) and critical limb ischemia (2.9% HRPCI, 38.5% CS).

CONCLUSION

The i-cor ECG-synchronized cardiac assist device appears safe and feasible showing clinical outcomes comparable to existing data in the setting of high-risk percutaneous coronary interventions and acute cardiogenic shock. Further prospective trials are warranted to identify optimal patient and interventional characteristics that will benefit most of this novel kind of mechanical circulatory support.

Для цитирования: Voigt Ingo, Spangenberg Tobias, Ibrahim Tareq, Bradaric Christian, Viertel Achim, Tallone Ezequiel Marcelo, Skurk Carsten, Abel Peter, Graf Jochen, Rinne Torsten, Böhm Jürgen, Ghanem Alexander, Liebetrau Christoph. Efficacy and safety of ECG-synchronized pulsatile extracorporeal membrane oxygenation in the clinical setting: The SynCor Trial.. Artificial organs. 2021. DOI: 10.1111/aor.14147.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.