DOI 10.1016/j.transproceed.2019.03.025
Comparison of the Outcome of Kidney Transplant After Pulsatile or Continuous Ex Vivo Hypothermic Machine Perfusion of Kidneys Donated After Cardiac Death: Analysis of Kidney Pairs.
Аннотация
BACKGROUND
Hypothermic machine perfusion is used to improve renal perfusion and reduce the rate of early and late graft dysfunction. It has been used in our unit since 2001. It has 2 modes of flow: continuous or pulsatile. The aim of this study is to compare the modes of perfusion in terms of perfusion-related parameters, graft survival, and estimated glomerular filtration rate.
METHODS
All donation after cardiac death kidneys between 2002 and 2014 were reviewed. A total of 64 pairs of kidneys were identified of which one kidney underwent pulsatile and the other continuous perfusion. Machine parameters including resistance and perfusion flow index levels at 0, 1, 2, 3, and 4 hours were recorded and glutathione S-transferase was measured in perfusate. Estimated glomerular filtration rate from the first week of transplant until the fifth year and graft survival rates were determined.
RESULTS
Machine parameters were similar at all time points. Estimated glomerular filtration rates and graft survival were the same irrespective of perfusion mode.
CONCLUSION
Pulsatile perfusion may be regarded as more physiological. However, we could not identify difference in outcome following transplant of kidneys from the same donor that had been perfused under pulsatile or continuous conditions.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.