False positive troponin result caused by a true macrotroponin.

Michielsen Etienne C H J, Bisschops Pieter G T, Janssen Marcel J W

Клинический случайСтатьяПеревод в очереди

Аннотация

BACKGROUND

A 35-year-old female presented with complaints of tiredness, shortness of breath and recent onset of chest pain. The cardiac troponin I (cTnI) concentration was elevated at 6.4 μg/L (Abbott Architect reference value: <0.03 μg/L) with a normal ECG. Physical examination, radiological imaging and routine laboratory investigations did not provide an explanation for the elevated cTnI concentration.

METHODS

Use of different troponin assays, dilutions with assay diluent, addition of mouse serum, polyethylene glycol (PEG) precipitation and gel filtration chromatography (GFC) were used to investigate possible interference.

RESULTS

Troponin concentrations were below the level of detection when measured using all other troponin immunoassays. Serial dilutions of sample and addition of mouse serum did not alter the results. However, PEG precipitation and GFC showed the presence of a high molecular weight immunoreactive protein. Using GFC and protein-A IgG precipitation, the interference could be identified as a macrocomplex containing IgG and (fragments of) cTnI.

CONCLUSIONS

We report a false positive cTnI result caused by a true macrotroponin, containing IgG and (fragments of) cTnI. This macrotroponin was only immunoreactive in the Abbott Architect cTnI immunoassay. Clinicians should be aware of analytical interference when troponin results are constantly elevated in the absence of (non)coronary causes of an increased troponin.

Для цитирования: Michielsen Etienne C H J, Bisschops Pieter G T, Janssen Marcel J W. False positive troponin result caused by a true macrotroponin.. Clinical chemistry and laboratory medicine. 2011. DOI: 10.1515/cclm.2011.147.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.