Inappropriate mode switching clarified by using a chest radiograph.
Аннотация
An 80-year-old woman with a history of paroxysmal atrial fibrillation and atrioventricular node disease status post-dual chamber pacemaker placement was noted to have abnormal pacing episodes during a percutaneous coronary intervention. Pacemaker interrogation revealed a high number of short duration mode switching episodes. Representative electrograms demonstrated high frequency nonphysiologic recordings predominantly in the atrial lead. Intrinsic pacemaker malfunction was excluded. A chest radiograph showed excess atrial and ventricular lead slack in the right ventricular inflow. It was suspected that lead-lead interaction resulted in artifacts and oversensing, causing frequent short episodes of inappropriate mode switching.