Work of breathing during nasal continuous positive airway pressure in preterm infants: a comparison of bubble vs variable-flow devices.

Liptsen Ellina, Aghai Zubair H, Pyon Kee H, Saslow Judy G, Nakhla Tarek, Long Jennifer, Steele Andrew M, Habib Robert H, Courtney Sherry E

Клиническое исследованиеСравнительное исследованиеСтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

To compare work of breathing and breathing asynchrony during bubble nasal continuous positive airway pressure (NCPAP) vs variable-flow (VF)-NCPAP in premature infants.

STUDY

DESIGN

We studied 18 premature infants of birth weight <1500 g who required NCPAP for mild respiratory distress. Each infant was studied on bubble and VF-NCPAP at 8, 6, 4, and 0 cm H2O. Tidal volumes were obtained by calibrated respiratory inductance plethysmography. Esophageal pressure estimated intrapleural pressure. Inspiratory and resistive work of breathing were calculated from pressure-volume data. Breathing asynchrony was assessed with phase angle. The results at all NCPAP levels were referenced to VF-NCPAP values at 8 cm H2O.

RESULTS

Provision of NCPAP with either device decreased inspiratory work of breathing, tidal volume, and minute ventilation relative to NCPAP of 0 cm H2O. Bubble NCPAP did not decrease resistive work of breathing relative to 0 cm H2O. Resistive work of breathing (p=0.01), respiratory rate (p<0.03), and phase angle (p=0.002) were all greater with bubble compared to VF-NCPAP.

CONCLUSION

The more labored and asynchronous breathing seen with bubble NCPAP may lead to higher failure rates over the long term than with VF-NCPAP.

Для цитирования: Liptsen Ellina, Aghai Zubair H, Pyon Kee H, Saslow Judy G, Nakhla Tarek, Long Jennifer, Steele Andrew M, Habib Robert H, Courtney Sherry E. Work of breathing during nasal continuous positive airway pressure in preterm infants: a comparison of bubble vs variable-flow devices.. Journal of perinatology : official journal of the California Perinatal Association. 2005. DOI: 10.1038/sj.jp.7211325.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.