Journal of perinatal medicine · 2005
Amniotic fluid lamellar body counts for the determination of fetal lung maturity: an update.
Клиническое исследованиеСтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Piazze Juan J, Maranghi Luca, Cerekja Albana, Meloni Paolo, Gioia Stefano, Fumian Luisa, Cosmi Ermelando V, Anceschi Maurizio M
Аннотация
AIM: To reassess the cut-off value for lamellar body counts (LBs) for fetal lung maturity (FLM) over a 10-year study period.
PATIENTS AND METHODS: 178 pregnancies were selected under strict inclusion criteria and delivered within 48 h from amniocentesis. FLM was determined by amniotic fluid LBs in centrifuged samples (300 x g for 10 min) in a commercially available Coulter Counter. Cases beyond 37 weeks were excluded.
RESULTS: Mean gestational age was 33.5+/-3.0 weeks at amniocentesis and 33.7+/-3.0 weeks at birth. After reassessing the best compromise between sensitivity and specificity for all cases using the receiver operating characteristic (ROC) procedure, an FLM cut-off value of < or = 22,000/microL was obtained. Diagnostic accuracy (and confidence interval, CI) was: sensitivity, 73% (60.0-83.6%); specificity, 81.7% (CI 73.6-88.1%); positive predictive value, 66.2%; and negative predictive value, 86.0%.
CONCLUSION: No significant change in FLM cut-off for LBs was found when comparing the value from this study and the results of our earlier report presented in 1996 (< or = 22,000 vs. < or = 20,000/microL), although the new value may be more accurate, since it is based on neonatal outcome with the exclusion of cases in which the diagnosis of FLM is seldom warranted, i.e., > 37 weeks' gestational age.
PATIENTS AND METHODS: 178 pregnancies were selected under strict inclusion criteria and delivered within 48 h from amniocentesis. FLM was determined by amniotic fluid LBs in centrifuged samples (300 x g for 10 min) in a commercially available Coulter Counter. Cases beyond 37 weeks were excluded.
RESULTS: Mean gestational age was 33.5+/-3.0 weeks at amniocentesis and 33.7+/-3.0 weeks at birth. After reassessing the best compromise between sensitivity and specificity for all cases using the receiver operating characteristic (ROC) procedure, an FLM cut-off value of < or = 22,000/microL was obtained. Diagnostic accuracy (and confidence interval, CI) was: sensitivity, 73% (60.0-83.6%); specificity, 81.7% (CI 73.6-88.1%); positive predictive value, 66.2%; and negative predictive value, 86.0%.
CONCLUSION: No significant change in FLM cut-off for LBs was found when comparing the value from this study and the results of our earlier report presented in 1996 (< or = 22,000 vs. < or = 20,000/microL), although the new value may be more accurate, since it is based on neonatal outcome with the exclusion of cases in which the diagnosis of FLM is seldom warranted, i.e., > 37 weeks' gestational age.