The EMCOOLs surface cooling system for fever control in neurocritical care patients: A pilot study.

Griffiths Stephen A, Ahmad Javaad, Francoeur Charles L, Gordon Errol, Dangayach Neha S, Wheelwright Danielle, Ramineni Anil, Mayer Stephan A

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

Fever occurs in up to 50% of critically-ill patients with acute neurological injury. Small temperature elevations have been correlated with increased morbidity and mortality in this patient population. We sought to evaluate a novel single-use surface cooling system for the treatment of fever in patients with acute brain injury.

PATIENTS AND

METHODS

We conducted a retrospective analysis of a prospective product evaluation using the EMCOOLS Flex.Pad™ system for acute fever (≥38.3 °C) in our 16-bed neuro-ICU. Four refrigerated pads (-18 °C) were applied to the chest, back, and anterior thighs. Core temperature (bladder) was continuously recorded over 4 h, and the highest Bedside Shivering Assessment Scale (BSAS) score was recorded hourly.

RESULTS

Twelve subjects were included in the analysis. Mean age was 55 ± 9 years, 9 patients were men, and mean weight was 85 ± 12 kg. The most common primary diagnoses were subarachnoid (N = 5) and intracerebral (N = 4) hemorrhage. Application of the EMCOOLS system resulted in a linear 1.3 ± 0.6 °C drop (T0avg = 38.9 ⁰C, T90avg = 37.6 ⁰C, P = 0.0032) in mean temperature over 90 min, followed by a plateau with only one subject rebounding to >38 °C within 4 h. Normothermia (<38.0 ⁰C) was achieved in all but one patient (92%) in an average of 65 min. Comatose patients displayed a non-significantly higher degree of cooling at 90 min than did awake subjects (ΔTcoma = 1.74 °C vs ΔTawake = 0.74 °C hr-1, P = 0.067). There was no observed skin irritation upon removal of the device for any patients.

CONCLUSION

The EMCOOLs system is a well-tolerated, safe and effective short-term intervention for control of fever in neurological patients. Future studies are needed to compare efficacy of the EMCOOLs to other devices and interventions.

Для цитирования: Griffiths Stephen A, Ahmad Javaad, Francoeur Charles L, Gordon Errol, Dangayach Neha S, Wheelwright Danielle, Ramineni Anil, Mayer Stephan A. The EMCOOLs surface cooling system for fever control in neurocritical care patients: A pilot study.. Clinical neurology and neurosurgery. 2019. DOI: 10.1016/j.clineuro.2019.105412.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.