Asleep Robot-Assisted Surgery for the Implantation of Subthalamic Electrodes Provides the Same Clinical Improvement and Therapeutic Window as Awake Surgery.

Lefranc Michel, Zouitina Yassine, Tir Mélissa, Merle Philippe, Ouendo Martial, Constans Jean-Marc, Godefroy Olivier, Peltier Johann, Krystkowiak Pierre

Сравнительное исследованиеОценочное исследованиеСтатьяПеревод в очереди

Аннотация

OBJECTIVE

To study the impact of not performing awake clinical evaluation during the robot-assisted implantation of subthalamic nucleus deep brain stimulation (STN-DBS) electrodes on the stimulation parameters and clinical outcomes in patients with Parkinson disease (PD).

METHODS

A total of 23 patients with PD underwent robot-assisted surgery for the bilateral implantation of STN-DBS electrodes. Thirteen patients received general anesthesia (GA) and a limited intraoperative evaluation (side effects only), and the other 10 patients received local anesthesia (LA) and a full evaluation. The primary endpoint was the therapeutic window (TW), defined as the difference between the mean voltage threshold for motor improvement and the mean voltage threshold for side effects in the active contacts at 12 months after surgery. Motor scores were measured as well.

RESULTS

The TW was similar in the LA and GA groups, with mean ± standard deviation values of 2.06 ± 0.53 V and 2.28 ± 0.99 V, respectively (P = 0.32). In the short term, the Unified Parkinson Disease Rating Scale (UPDRS) III score in the "off-drug, on-stim" condition fell to a similar extent in the LA and GA groups (by 40.3% and 49%, respectively; P = 0.336), as did the UPDRS III score in the "on-stim, on-drug" condition (by 57% and 70.7%, respectively; P = 0.36).

CONCLUSIONS

Asleep, robot-assisted implantation of STN-DBS electrodes (with accurate identification of the STN and positioning of the DBS lead) produced the same motor results and TW as awake surgery.

Для цитирования: Lefranc Michel, Zouitina Yassine, Tir Mélissa, Merle Philippe, Ouendo Martial, Constans Jean-Marc, Godefroy Olivier, Peltier Johann, Krystkowiak Pierre. Asleep Robot-Assisted Surgery for the Implantation of Subthalamic Electrodes Provides the Same Clinical Improvement and Therapeutic Window as Awake Surgery.. World neurosurgery. 2017. DOI: 10.1016/j.wneu.2017.07.047.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.