DOI 10.1111/ijcp.12856

Медицинская библиотека · International journal of clinical practice · 2016

Feasibility of spectral imaging with low-concentration contrast medium in abdominal CT angiography of obese patients.

Liu Yijun, Liu Ailian, Liu Lei, Tian Shifeng, Liu Jinghong, Pu Renwang, Fang Xin, Liu Xiaofeng, Yuan Gang

Клиническое исследованиеСравнительное исследованиеСтатьяПеревод в очереди

Аннотация

OBJECTIVE

To explore the application value of dual-energy spectral imaging scanning with low radiation dose and low-concentration contrast medium (270 mg I/mL) in abdominal CT angiography (CTA) of obese patients.

METHODS

A total of 127 obese patients (BMI≥25 kg/m(2) and waist circumference ≥900 mm) referred for abdominal CTA were prospectively enrolled in the study. The patients were divided into two groups; in group A (n=69), a spectral imaging scan mode and 270 mg I/mL iodine concentration contrast medium was used, and in group B (n=58), a conventional imaging scan mode using 120 kVp and 350 mg I/mL iodine concentration contrast medium was used. The image quality of the right renal artery in the two groups was evaluated by two observers using a 5-point scale, and the scores were compared using the Mann-Whitney U-test; the inter-observer agreement for the scores was analysed using the Kappa test. The CT values of the abdominal aorta, the superior mesenteric artery, the common hepatic artery and the splenic artery, and the CT value and standard deviation (SD) of the erector spinae at the level of the right renal hilum in groups A and B were measured by two observers; the inter-observer agreement of the measurement data was analysed using the inter-class correlation coefficient test. The following parameters were compared between the two groups using an independent sample t-test: the CT values of the abdominal aorta and its main branches; the image contrast-to-noise ratio (CNR) and figure of merit (FOM) of the abdominal aorta; the CT dose index (CTDIvol ); the dose length product (DLP); and the total iodine intake of the patients. P<.05 suggested a statistically significant difference.

RESULTS

The image scores of the right renal artery in groups A (4.59±0.60) and B (4.53±0.63) were the same (P=.57), with good inter-observer agreement. The CT values of the abdominal aorta, the superior mesenteric artery, the common hepatic artery and the splenic artery were >300 HU in both the groups; there was no statistically significant difference between the two groups (all P>.05), and inter-observer agreement was also good. Group A had significantly higher CNR and FOM values in the abdominal aorta than group B (all P<.001). Compared with group B, the CTDIvol and DLP values in group A were decreased by 46% and 35%, respectively. The total iodine intake for patients in group A was 27 g, 23% lower than the 35 g intake for patients in group B.

CONCLUSION

Compared with conventional CT, spectral CT imaging significantly reduces both radiation dose and contrast dose while maintaining image quality in abdominal CTA for patients with central obesity and high BMI.

Для цитирования: Liu Yijun, Liu Ailian, Liu Lei, Tian Shifeng, Liu Jinghong, Pu Renwang, Fang Xin, Liu Xiaofeng, Yuan Gang. Feasibility of spectral imaging with low-concentration contrast medium in abdominal CT angiography of obese patients.. International journal of clinical practice. 2016. DOI: 10.1111/ijcp.12856.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.