DOI 10.1016/0378-4347(96)00138-7
Separation of neutral lipids by high-performance liquid chromatography: quantification by ultraviolet, light scattering and fluorescence detection.
Аннотация
The recent increased use of cell cultures to model physiological events, in particular signal transduction and traumatic injury, has produced a need for a quantitative, high-performance liquid chromatographic separation of neutral lipid classes with a high degree of resolution and reproducibility. We report an isocratic separation using a Phenomenex Selectosil silica column (5 microns). Two solvents were used, 1.2% 2-propanol in n-hexane containing 0.1% acetic acid (90%) and n-hexane (10%) at a flow-rate of 0.6 ml/min. Column temperature was maintained at 55 degrees C and this temperature was critical for baseline resolution of 1,3-diacylglycerol and cholesterol. The use of 10% n-hexane permitted the resolution of low polarity compounds such as butylated hydroxytoluene, triacylglycerols and cholesteryl esters. All of the detectors used produced standard curves with linearity over a wide concentration range.