Anesthetic breathing circuit obstruction mimicking severe bronchospasm: an unusual manufacturing defect.

Yang Chien-Hui, Chen Kuan-Hung, Lee Ying-En, Lin Chung-Ren

Клинический случайСтатьяПеревод в очереди

Аннотация

We report an unexpected ventilation difficulty with an anesthetic breathing circuit in a pediatric patient receiving left herniorrhaphy. A manufacturing defect in a limb of the anesthetic breathing circuit caused this problem. This defect induced a high-pitched, wheezing-like sound, which was difficult to differentiate from a hyper-reactive airway, commonly seen in pediatric patients with recent upper respiratory tract infection. We recommend that the patency of the anesthetic breathing circuit should routinely be examined before connecting it to the anesthesia machine.

Для цитирования: Yang Chien-Hui, Chen Kuan-Hung, Lee Ying-En, Lin Chung-Ren. Anesthetic breathing circuit obstruction mimicking severe bronchospasm: an unusual manufacturing defect.. Acta anaesthesiologica Taiwanica : official journal of the Taiwan Society of Anesthesiologists. 2012. DOI: 10.1016/j.aat.2012.02.005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.