Intracardiac echocardiography evaluation in secundum atrial septal defect transcatheter closure.

Zanchetta Mario, Pedon Luigi, Rigatelli Gianluca, Carrozza Antonio, Zennaro Marco, Di Martino Roberta, Onorato Eustaquio, Maiolino Pietro

СтатьяПеревод в очереди

Аннотация

PURPOSE: This study was designed to assess the balloon sizing maneuvers and deployment of an Amplatzer Septal Occluder (ASO). In addition, intraprocedural balloon sizing was compared with off-line intracardiac echocardiographic measurements.

METHODS

The intracardiac echocardiography (ICE) measurements were: maximum transverse and longitudinal atrial septal defect (ASD) diameters in the aortic valve and four-chamber planes; area of the ASD and its equivalent circle diameter. Thirteen consecutive patients underwent transcatheter implantation of an ASO device using ICE guidance under local anesthesia. The device matching the balloon sizing diameter of the defect was implanted. Qualitative ICE assessment of the ASO devices implanted was performed off line.

RESULTS

The mean equivalent circle diameter predicted by ICE was 24.40 +/- 5.61 mm and was significantly higher (p = 0.027) than the ASD measured by balloon sizing (21.38 +/- 5.28 mm). Unlike previous studies we did not find any correlation between the two measurements (correlation coefficient = 0.47). Only four of the 13 patients had optimal device positioning as shown by the qualitative ICE evaluation, whereas the remaining nine patients had inadequate device placement. This resulted in a waist diameter that was an average 26.1% undersized in seven patients and 12.7% oversized in two patients. Five of the seven patients with an undersized device had ASO-atrial septum misalignment with leftward device deviation.

CONCLUSION

The ICE images allowed careful measurement of the dimensions of the ASD and accurately displayed the spatial relations of the ASO astride the ASD. Moreover, use of the ICE measurement led to selection of a different size of device in comparison with those of balloon sizing. The clinical benefit of this new approach needs to be rigorously tested.

Для цитирования: Zanchetta Mario, Pedon Luigi, Rigatelli Gianluca, Carrozza Antonio, Zennaro Marco, Di Martino Roberta, Onorato Eustaquio, Maiolino Pietro. Intracardiac echocardiography evaluation in secundum atrial septal defect transcatheter closure.. Cardiovascular and interventional radiology. 2002. DOI: 10.1007/s00270-002-1928-7.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.