DOI 10.1093/chromsci/bmae021

Медицинская библиотека · Journal of chromatographic science · 2025

Development and Validation of Fast and Sensitive RP-HPLC Stability-Indicating Method for Quantification of Piroxicam in Bulk Drug.

Chauhan Vikas, Grover Parul, Bhardwaj Monika, Kumar Sandeep, Nagarajan K

СтатьяПеревод в очереди

Аннотация

A simple, rapid, sensitive, and cost-effective green solvent-assisted reverse-phase high-performance liquid chromatographic technique, coupled with a photodiode array detector, was developed and validated for the estimation of piroxicam (PRXM). The chromatographic separation was achieved by using a C-18 (250 × 4.6) mm, 5-μm stationary phase and a mobile phase consisting of methanol and 0.1% ortho-phosphoric acid in water in a ratio of (80:20) v/v at a flow rate of 1 ml/min. The detection was carried out at a wavelength of 254 nm with a constant injection volume of 10 μL throughout the analysis. The calibration curve was observed to be linear over the optimum concentration range of 50-300 μg mL-1, with an R2 value of 0.9995. The developed method was validated as per the International Council for Harmonisation (ICH) Q2 (R1) guideline. Various parameters like selectivity/specificity, accuracy/recovery, linearity, precision, detection limit, quantitation limit, robustness and stability of analyte in solution were performed for the method validation. The PRXM was evaluated under stressed conditions, including acidic, basic, oxidative, thermal and photolytic, as per ICH Q1 (R2) guidelines. Significant degradation was observed in acidic and basic degradation conditions. Conversely, the drug substance showed stability when exposed to oxidative, photolytic and thermal degradation conditions.

Для цитирования: Chauhan Vikas, Grover Parul, Bhardwaj Monika, Kumar Sandeep, Nagarajan K. Development and Validation of Fast and Sensitive RP-HPLC Stability-Indicating Method for Quantification of Piroxicam in Bulk Drug.. Journal of chromatographic science. 2025. DOI: 10.1093/chromsci/bmae021.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.