OBJECTIVES

The main goal of this study was to compare the occurrence of post-deployment leaflet injury between prostheses made of porcine and bovine pericardium.

METHODS

Two types of prostheses, self-expandable prostheses with porcine pericardial leaflet on one side and balloon-expandable prostheses with bovine pericardial leaflet on the other side, were used. In each group, crimped prostheses were compared with control non-crimped prostheses. Following a 15-min period of crimping, prostheses were deployed, and their leaflets were removed and subjected to analyses. The analyses included determination of global and local hydraulic conductance of the leaflets, leaflet plasma insudation and microscopic analysis. The results were expressed as percentage (standard error of the mean) or median (interquartile range).

RESULTS

A significant increase in the global hydraulic conductance was only observed in the crimped balloon-expandable prostheses: 20.1 (15.5-41.2) ml/h/m2/mmHg vs 12.3 (9.0-15.6) ml/h/m2/mmHg (P = 0.021). Similarly, areas of marked staining (a marker of local hydraulic conductance) were only seen in the bovine pericardium balloon-expandable prostheses. The incidence of leaflet plasmatic insudation was increased in the crimped prostheses compared with the control prostheses. The microscopic study revealed a higher occurrence of traumatic lesions in the crimped prostheses in comparison with the control prostheses: 33.3 ± 21.1% vs 5.5 ± 5.5% (P = 0.039) and 44.4 ± 20.5% vs 5.5 ± 5.5% (P = 0.017) in the bovine and the porcine leaflets, respectively.

CONCLUSIONS

Post-deployment leaflet injury occurred in both types of prostheses. However, alteration of the global and local hydraulic conductance was important in the bovine pericardium balloon-expandable prostheses.

Для цитирования: Convelbo Channing, El Hafci Hanane, Petite Hervé, Zegdi Rachid. Traumatic leaflet injury: comparison of porcine leaflet self-expandable and bovine leaflet balloon-expandable prostheses.. European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery. 2018. DOI: 10.1093/ejcts/ezx451.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.