Analytical evaluation of a direct ion-selective-based analyser: Still gaps to close.

Oyaert Matthijs, Verougstraete Nick, Vandekerckhove Brecht, Lapauw Bruno, Hoste Eric, Stove Veronique

СтатьяОценочное исследованиеПеревод в очереди

Аннотация

INTRODUCTION

Discrepancies between electrolyte concentrations determined by blood gas analysers (BGA) and core-lab chemistry analysers may create confusion in clinical practice. This problem is rooted in the different ion-selective electrode (ISE) methodologies that are used. Whilst most available chemistry analysers use indirect ISE, we evaluated the analytical performance of the new automated chemistry analyser Biossays™ E6 (Snibe), equipped with direct ISE, for the determination of sodium (Na+), potassium (K+), chloride (Cl-), ionized calcium (iCa2+) and pH.

MATERIALS AND

METHODS

Total precision, estimated deviation and total error were evaluated for all analytes on the E6 analyser. Several patient cohorts were used to perform method comparisons between the E6 and the direct (RP500e BGA) and indirect (Architect c16000 analyser) ISE methods routinely used in the lab. Obtained data were compared against pre-set quality specifications and used for adjustment of the 2 direct ISE methods. For Na+ and iCa2+, agreement with the routinely used protein-corrected Na+ and total calcium (TCa2+) concentrations were assessed respectively.

RESULTS

The analytical performance for the 4 tested electrolytes (Na+, K+, Cl-, iCa2+) and pH were acceptable and within the specified performance specifications. After adjustment of both direct methods, method comparison on an independent patient cohort showed good agreement. For Na+ and iCa2+, a good correlation with the protein corrected Na+ and TCa2+ results was observed.

CONCLUSION

The acceptable analytical performance and ease-of-use of the E6 direct ion selective instrument is making it feasible to optimize electrolyte determinations to direct methodology.

Для цитирования: Oyaert Matthijs, Verougstraete Nick, Vandekerckhove Brecht, Lapauw Bruno, Hoste Eric, Stove Veronique. Analytical evaluation of a direct ion-selective-based analyser: Still gaps to close.. Clinical biochemistry. 2024. DOI: 10.1016/j.clinbiochem.2024.110829.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.