DOI 10.1007/s00428-020-02962-x

Медицинская библиотека · Virchows Archiv : an international journal of pathology · 2020

Multi-center real-world comparison of the fully automated Idylla™ microsatellite instability assay with routine molecular methods and immunohistochemistry on formalin-fixed paraffin-embedded tissue of colorectal cancer.

Velasco Ana, Tokat Fatma, Bonde Jesper, Trim Nicola, Bauer Elisabeth, Meeney Adam, de Leng Wendy, Chong George, Dalstein Véronique, Kis Lorand L, Lorentzen Jon A, Tomić Snjezana, Thwaites Keeley, Putzová Martina, Birnbaum Astrid, Qazi Romena, Primmer Vanessa, Dockhorn-Dworniczak Barbara, Hernández-Losa Javier, Soares Fernando A, Gertler Asaf A, Kalman Michal, Wong Chris, Carraro Dirce M, Sousa Ana C, Reis Rui M, Fox Stephen B, Fassan Matteo, Brevet Marie, Merkelbach-Bruse Sabine, Colling Richard, Soilleux Elizabeth, Teo Ryan Yee Wei, D'Haene Nicky, Nolet Serge, Ristimäki Ari, Väisänen Timo, Chapusot Caroline, Soruri Afsaneh, Unger Tina, Wecgowiec Johanna, Biscuola Michele, Frattini Milo, Long Anna, Campregher Paulo V, Matias-Guiu Xavier

Сравнительное исследованиеСтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

Microsatellite instability (MSI) is present in 15-20% of primary colorectal cancers. MSI status is assessed to detect Lynch syndrome, guide adjuvant chemotherapy, determine prognosis, and use as a companion test for checkpoint blockade inhibitors. Traditionally, MSI status is determined by immunohistochemistry or molecular methods. The Idylla™ MSI Assay is a fully automated molecular method (including automated result interpretation), using seven novel MSI biomarkers (ACVR2A, BTBD7, DIDO1, MRE11, RYR3, SEC31A, SULF2) and not requiring matched normal tissue. In this real-world global study, 44 clinical centers performed Idylla™ testing on a total of 1301 archived colorectal cancer formalin-fixed, paraffin-embedded (FFPE) tissue sections and compared Idylla™ results against available results from routine diagnostic testing in those sites. MSI mutations detected with the Idylla™ MSI Assay were equally distributed over the seven biomarkers, and 84.48% of the MSI-high samples had ≥ 5 mutated biomarkers, while 98.25% of the microsatellite-stable samples had zero mutated biomarkers. The concordance level between the Idylla™ MSI Assay and immunohistochemistry was 96.39% (988/1025); 17/37 discordant samples were found to be concordant when a third method was used. Compared with routine molecular methods, the concordance level was 98.01% (789/805); third-method analysis found concordance for 8/16 discordant samples. The failure rate of the Idylla™ MSI Assay (0.23%; 3/1301) was lower than that of referenced immunohistochemistry (4.37%; 47/1075) or molecular assays (0.86%; 7/812). In conclusion, lower failure rates and high concordance levels were found between the Idylla™ MSI Assay and routine tests.

Для цитирования: Velasco Ana, Tokat Fatma, Bonde Jesper, Trim Nicola, Bauer Elisabeth, Meeney Adam, de Leng Wendy, Chong George, Dalstein Véronique, Kis Lorand L, Lorentzen Jon A, Tomić Snjezana, Thwaites Keeley, Putzová Martina, Birnbaum Astrid, Qazi Romena, Primmer Vanessa, Dockhorn-Dworniczak Barbara, Hernández-Losa Javier, Soares Fernando A, Gertler Asaf A, Kalman Michal, Wong Chris, Carraro Dirce M, Sousa Ana C, Reis Rui M, Fox Stephen B, Fassan Matteo, Brevet Marie, Merkelbach-Bruse Sabine, Colling Richard, Soilleux Elizabeth, Teo Ryan Yee Wei, D'Haene Nicky, Nolet Serge, Ristimäki Ari, Väisänen Timo, Chapusot Caroline, Soruri Afsaneh, Unger Tina, Wecgowiec Johanna, Biscuola Michele, Frattini Milo, Long Anna, Campregher Paulo V, Matias-Guiu Xavier. Multi-center real-world comparison of the fully automated Idylla™ microsatellite instability assay with routine molecular methods and immunohistochemistry on formalin-fixed paraffin-embedded tissue of colorectal cancer.. Virchows Archiv : an international journal of pathology. 2020. DOI: 10.1007/s00428-020-02962-x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.