Single-Use Bag Valve Masks: Evaluation of Device Design and Residual Bioburden Analytical Methods.

Zemitis Sarah, Harman Melinda, Hargett Zachary, Weinbrenner Donna

СтатьяПеревод в очереди

Аннотация

BACKGROUND

A recent survey of in-hospital reprocessing in Tanzanian hospitals identified bag-valve masks (BVM) as a commonly reused single-use device. In low- and middle-income countries (LMIC), in-hospital reprocessing supports neonatal resuscitation strategies by helping to maintain adequate supplies of BVM. However, there is a need for device-specific protocols defining reprocessing procedures and inspection criteria to overcome variations in reprocessing practices between hospitals. The purposes of this study were: 1) to complete a comprehensive design review and identify challenges to reprocessing BVMs; and 2) to investigate three different residual bioburden analysis methods for assessing the efficacy of decontaminating a disposable BVM.

METHODS

New, unused bag-valve-masks were contaminated with Staphylococcus epidermidis and Artificial Mucus Soil to simulate the worst case soiling conditions. Devices underwent one of five disinfection protocols, including one currently used in a LMIC hospital. Three analytical (two quantitative and one qualitative) methods were selected to evaluate residual bioburden on the device following decontamination.

RESULTS

Of all protocols tested, only the positive control and the Soap and Bleach protocols met disinfection targets. Most cleaning outcomes were consistent from trial to trial for each protocol. However, cleaning outcomes varied greatly for the Alcohol Wipe protocol. For the residual bioburden analyses, the two quantitative methods produced similar results, but the qualitative measurement exhibited increased variability.

CONCLUSION

While this study revealed positive disinfection outcomes for the Tanzanian hospital decontamination protocol, more studies are required to support these findings. Design features of the BVM mask presented challenges to cleaning and drying during different decontamination protocols, as seen in the variability in the Alcohol Wipe protocol performance. These findings support the case for a device-specific protocol for the BVM. Given proper hospital personnel training and available resources, in-hospital reprocessing could support neonatal resuscitation strategies and other demands for manual resuscitation by helping to maintain adequate supplies of BVM.

Для цитирования: Zemitis Sarah, Harman Melinda, Hargett Zachary, Weinbrenner Donna. Single-Use Bag Valve Masks: Evaluation of Device Design and Residual Bioburden Analytical Methods.. Journal of biomedical science and engineering. 2018. DOI: 10.4236/jbise.2018.119019.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.