Fractured inlet connecting tube of the flat bottom flushing device of a posterior fossa cystoperitoneal shunt.

Amirjamshidi Abbas, Abbasioun Kazem, Ghassemi Babak

Клинический случайСтатьяПеревод в очереди

Аннотация

BACKGROUND

There are well-known complications for shunt procedures. Shunt fracture or disconnection is the second most frequent cause of shunt malfunction in children. Shunt disconnection is not a common cause of shunt malfunction in the early period after installation, especially in the adulthood.

CASE DESCRIPTION: Fracture of the proximal (inlet) connector of a flat-based shunt installed for decompression of a large posterior fossa arachnoid cyst in a 31-year-old female with signs of increased intracranial pressure led to recurrence of her symptoms 6 months after surgery.

CONCLUSION

Awareness of the possibility of fracture site in the junction of the inlet connector of flat bottom shunt systems is warranted and can be diagnosed by three-dimensional computed tomography (3D CT) imaging without performing shunt series study.

Для цитирования: Amirjamshidi Abbas, Abbasioun Kazem, Ghassemi Babak. Fractured inlet connecting tube of the flat bottom flushing device of a posterior fossa cystoperitoneal shunt.. Surgical neurology international. 2015. DOI: 10.4103/2152-7806.152734.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.