Chu Michael W A, Bagur Rodrigo, Losenno Katie L, Jones Philip M, Diamantouros Pantelis, Teefy Patrick, Gelinas Jill J, Kiaii Bob
Аннотация
OBJECTIVE: Coronary obstruction remains a challenging complication of transcatheter aortic valve replacement; however, a new self-expanding transapical prosthesis may reduce this risk. The purpose of this study was to evaluate the early 1-year outcomes of patients with low coronary heights who received the Acurate TA bioprosthesis (Symetis, Ecublens, Switzerland).
METHODS: Between May 2014 and April 2015, 30 consecutive patients (aged 85 ± 6 years, 63% were female, Society of Thoracic Surgeons score 8.4 ± 6.0) with severe, symptomatic aortic stenosis underwent transcatheter aortic valve replacement with the Acurate TA bioprosthesis. Relevant patient characteristics included reoperation in 47% (n = 14), peripheral vascular disease in 43% (n = 13), and porcelain aorta in 30% (n = 9). The mean left and right coronary heights were 10.8 ± 1.5 mm and 16.4 ± 4.1 mm, respectively, with a sinus of Valsalva : annular ratio of 1.3 ± 0.8.
RESULTS: All 30 device implants were successful. The 30-day in-hospital mortality was 3.3% (n = 1), and no patients had coronary obstruction or stroke. One patient (3.3%) had apical rupture requiring cardiopulmonary bypass for repair, 1 patient (3.3%) had a localized femoral artery dissection, and 1 patient (3.7%) required a new pacemaker. There were no other complications. Mean and peak transaortic valve gradients decreased from 59 ± 17 and 84 ± 31 mm Hg to 14 ± 7 and 28 ± 12 mm Hg, respectively (P < .0005). No patients had more than mild paravalvular aortic insufficiency. At 30 days, there were no further complications and 96.7% (n = 29) were in New York Heart Association class I/II. Survival at 30 days and 1 year was 97% and 89%, respectively.
CONCLUSIONS: The Symetis Acurate TA device demonstrates high procedural success and excellent acute and 1-year patient outcomes. The device allows safe implantation in patients at higher risk for coronary artery obstruction.
METHODS: Between May 2014 and April 2015, 30 consecutive patients (aged 85 ± 6 years, 63% were female, Society of Thoracic Surgeons score 8.4 ± 6.0) with severe, symptomatic aortic stenosis underwent transcatheter aortic valve replacement with the Acurate TA bioprosthesis. Relevant patient characteristics included reoperation in 47% (n = 14), peripheral vascular disease in 43% (n = 13), and porcelain aorta in 30% (n = 9). The mean left and right coronary heights were 10.8 ± 1.5 mm and 16.4 ± 4.1 mm, respectively, with a sinus of Valsalva : annular ratio of 1.3 ± 0.8.
RESULTS: All 30 device implants were successful. The 30-day in-hospital mortality was 3.3% (n = 1), and no patients had coronary obstruction or stroke. One patient (3.3%) had apical rupture requiring cardiopulmonary bypass for repair, 1 patient (3.3%) had a localized femoral artery dissection, and 1 patient (3.7%) required a new pacemaker. There were no other complications. Mean and peak transaortic valve gradients decreased from 59 ± 17 and 84 ± 31 mm Hg to 14 ± 7 and 28 ± 12 mm Hg, respectively (P < .0005). No patients had more than mild paravalvular aortic insufficiency. At 30 days, there were no further complications and 96.7% (n = 29) were in New York Heart Association class I/II. Survival at 30 days and 1 year was 97% and 89%, respectively.
CONCLUSIONS: The Symetis Acurate TA device demonstrates high procedural success and excellent acute and 1-year patient outcomes. The device allows safe implantation in patients at higher risk for coronary artery obstruction.
Предметные рубрики MeSH
AgedAged, 80 and overAortic ValveAortic Valve StenosisBalloon ValvuloplastyBioprosthesisFemaleHeart Valve ProsthesisHumansMalePostoperative ComplicationsProsthesis DesignRecovery of FunctionRetrospective StudiesRisk FactorsSeverity of Illness IndexTime FactorsTranscatheter Aortic Valve ReplacementTreatment Outcome