PURPOSE: Percutaneous mechanical circulatory support (pMCS) with the Impella® device is routinely used in cardiogenic shock and high risk percutaneous coronary interventions (PCI). However, safety and feasibility to close the large bore access (LBA) post-hoc remain a challenge.

MATERIALS AND

METHODS

In patients with planned termination of Impella therapy, corresponding LBA closure was attempted using an undersized AngioSeal® device, which was deployed after insertion of a 0.035 in J-wire through the guidewire access port. Cross-over angiographic visualization before and after deployment as well as in-hospital follow up were performed to assess closure success and screening for short-term adverse events.

RESULTS

We evaluated 17 patients (68 year old, IQR 58-76, 71% male) in whom 14 F LBA closure after pMCS using the Impella CP® was achieved with an undersized AngioSeal® device. Clinical indication for pMCS was cardiogenic shock in 94% and protected PCI in 6%. Impella CP® devices were withdrawn after a median of 4 days (IQR 3-6 days). Primary hemostasis was achieved in the majority of patients (14/17; 82%) while single cross-over balloon inflation led to hemostasis in the remaining patients. One patient suffered acute ischemia later in the course of the disease, but this was not directly related to the closure device. In this study, no major bleeding or other device-related adverse events were observed.

CONCLUSION

The 8 F AngioSeal® vascular closure device has been safely used for removal of the Impella CP® microaxial pump in most cases of predominantly normal weight patients. This technique may be considered feasible for post-hoc LBA closure in the context of pMCS, especially when preclosure options are not suitable or unavailable.

Для цитирования: Zachoval Christian F, Sedaghat Alexander, Oezkur Mehmet, Tiyerili Vedat, Becher Marc U, Sinning Jan-Malte, Treede Hendrik, Nickenig Georg, Zimmer Sebastian. Feasibility and Safety of an 8 F Angioseal® Vascular Closure Device for Closure of Large Bore Impella CP® Access.. Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists. 2021. DOI: 10.1177/15266028211047965.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.