DOI 10.2460/ajvr.21.11.0192

Медицинская библиотека · American journal of veterinary research · 2022

Point-of-care ketone meters may be used to estimate serum β-hydroxybutyrate concentrations in healthy African penguins (Spheniscus demersus).

Meade Sharonda M, Di Girolamo Nicola, Cray Carolyn, Aquilino Nicolette, DiGeronimo Peter M, Brandão João

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To evaluate the agreement between 3 point-of-care (POC) devices and a reference laboratory for measuring β-hydroxybutyrate (β-HB) concentration in African penguin (Spheniscus demersus) whole blood (WB) and plasma samples and the precision of each POC device for measuring β-HB concentration in plasma samples.

ANIMALS: 48 healthy African penguins.

PROCEDURES: Blood was obtained from the right jugular vein of each penguin, and β-HB concentration was measured on each POC device using fresh WB and heparinized plasma and at the reference laboratory using plasma. β-HB concentration was measured in plasma on each POC device.

RESULTS

All devices overestimated serum β-HB concentrations on average by 0.46 mM relative to the reference laboratory. WB samples had less error than plasma for meters A and C. Meter A had the lowest total error observed (26.4%) and the lowest mean difference (0.19 mmol/L) relative to the reference laboratory. Controlling for other factors, the magnitude of disagreement was not affected by sex, age, packed cell volume, or serum total solids concentration.

CLINICAL RELEVANCE: WB, not plasma, should be used for measurement of β-HB concentration on the POC meters tested. Meter A showed good correlation with the reference laboratory for WB. The use of POC devices for the measurement of β-HB concentration may be acceptable when laboratory analyzers are not available. Further research is needed for clinical application and the diagnostic value of POC meters compared with reference laboratories.

Для цитирования: Meade Sharonda M, Di Girolamo Nicola, Cray Carolyn, Aquilino Nicolette, DiGeronimo Peter M, Brandão João. Point-of-care ketone meters may be used to estimate serum β-hydroxybutyrate concentrations in healthy African penguins (Spheniscus demersus).. American journal of veterinary research. 2022. DOI: 10.2460/ajvr.21.11.0192.