Performance evaluation of platelet counting by novel fluorescent dye staining in the XN-series automated hematology analyzers.

Tanaka Yuzo, Tanaka Yumiko, Gondo Kazumi, Maruki Yoshiko, Kondo Tamiaki, Asai Satomi, Matsushita Hiromichi, Miyachi Hayato

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Conventional automated hematology analyzers have limitations in platelet measurements such as poor accuracy and precision in the low count range and interference by nonplatelet particles. In order to improve it, the newly developed XN-Series automated hematology analyzers (Sysmex Corporation, Kobe, Japan) have been installed with a new dedicated channel for platelet analysis (PLT-F), which is based on a fluorescence flow cytometry method with uses of a novel fluorescent dye specifically staining platelets. We evaluated the basic performance of this new PLT-F channel.

METHODS

Basic performance of the PLT-F channel in within-run reproducibility and assay linearity was studied using standard methods. Correlation was studied between PLT-F and a conventional automated hematology analyzer (XE-2100) and immunoplatelet analysis using anti-CD61 monoclonal antibody (Cell-Dyn Sapphire; Abbott Laboratories). The assay interference by nonplatelet particles such as fragmented red and white blood cells was evaluated by using clinical samples, respectively, from burn injury and acute leukemia.

RESULTS

Basic performance of the PLT-F platelet counting was satisfactory in within-run reproducibility, linearity and correlation with the conventional analyzer. The correlation was satisfactory also with the immunoplatelet analysis, even for samples from a patient with burn injury, and those with white blood cell fragments displayed, platelet abnormal flag and low platelet counts (<50 × 10(9)/l).

CONCLUSION

The platelet counting performance of the PLT-F channel of the XN Series had improved accuracy and precision in the low range and in abnormal samples, avoiding the interference by nonplatelet particles.

Для цитирования: Tanaka Yuzo, Tanaka Yumiko, Gondo Kazumi, Maruki Yoshiko, Kondo Tamiaki, Asai Satomi, Matsushita Hiromichi, Miyachi Hayato. Performance evaluation of platelet counting by novel fluorescent dye staining in the XN-series automated hematology analyzers.. Journal of clinical laboratory analysis. 2014. DOI: 10.1002/jcla.21691.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.