Медицинская библиотека · Operative neurosurgery (Hagerstown, Md.) · 2021

Clinical Accuracy, Technical Precision, and Workflow of the First in Human Use of an Augmented-Reality Head-Mounted Display Stereotactic Navigation System for Spine Surgery.

Molina Camilo A, Sciubba Daniel M, Greenberg Jacob K, Khan Majid, Witham Timothy

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Augmented reality mediated spine surgery is a novel technology for spine navigation. Benchmark cadaveric data have demonstrated high accuracy and precision leading to recent regulatory approval. Absence of respiratory motion in cadaveric studies may positively bias precision and accuracy results and analogous investigations are prudent in live clinical scenarios.

OBJECTIVE

To report a technical note, accuracy, precision analysis of the first in-human deployment of this technology.

METHODS

A 78-yr-old female underwent an L4-S1 decompression, pedicle screw, and rod fixation for degenerative spine disease. Six pedicle screws were inserted via AR-HMD (xvision; Augmedics, Chicago, Illinois) navigation. Intraoperative computed tomography was used for navigation registration as well as implant accuracy and precision assessment. Clinical accuracy was graded per the Gertzbein-Robbins (GS) scale by an independent neuroradiologist. Technical precision was analyzed by comparing 3-dimensional (3D) (x, y, z) virtual implant vs real implant position coordinates and reported as linear (mm) and angular (°) deviation. Present data were compared to benchmark cadaveric data.

RESULTS

Clinical accuracy (per the GS grading scale) was 100%. Technical precision analysis yielded a mean linear deviation of 2.07 mm (95% CI: 1.62-2.52 mm) and angular deviation of 2.41° (95% CI: 1.57-3.25°). In comparison to prior cadaveric data (99.1%, 2.03 ± 0.99 mm, 1.41 ± 0.61°; GS accuracy 3D linear and angular deviation, respectively), the present results were not significantly different (P > .05).

CONCLUSION

The first in human deployment of the single Food and Drug Administration approved AR-HMD stereotactic spine navigation platform demonstrated clinical accuracy and technical precision of inserted hardware comparable to previously acquired cadaveric studies.

Для цитирования: Molina Camilo A, Sciubba Daniel M, Greenberg Jacob K, Khan Majid, Witham Timothy. Clinical Accuracy, Technical Precision, and Workflow of the First in Human Use of an Augmented-Reality Head-Mounted Display Stereotactic Navigation System for Spine Surgery.. Operative neurosurgery (Hagerstown, Md.). 2021. DOI: 10.1093/ons/opaa398.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.