DOI 10.1016/0021-9673(92)85009-i
Determination of tetracyclines in bovine and porcine muscle by high-performance liquid chromatography using solid-phase extraction.
Аннотация
A method is presented for the determination of the three tetracyclines oxytetracycline, tetracycline and chlortetracycline in muscle, spiked at 100 ng/g, using high-performance liquid chromatography (HPLC). The concentration and extraction steps are carried out using Waters Environmental Sep-Pak cartridges. The principal steps involve homogenizing the sample in EDTA-McIlvaine buffer followed by centrifugation and precipitation of the supernatant using trichloroacetic acid. After further filtration and concentration on a Sep-Pak cartridge, the sample is eluted and analysed by HPLC with UV detection and confirmation by diode-array. The column used is a Nova-Pak C18 (4 microns) cartridge (10 cm x 8 mm I.D.). A phosphate-citrate-acetonitrile buffer, utilizing ion suppression, is the mobile phase. The analytes are detectable at levels down to 10 ng/g. The analyte identity can be confirmed at 20 ng/g by the use of diode-array detection and spectral library comparison.