Rapid communications in mass spectrometry : RCM · 2015

Rapid analysis of forensic samples using an atmospheric solid analysis probe interfaced to a linear ion trap mass spectrometer.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Jagerdeo Eshwar, Clark Jay A, Leibowitz Jeffrey N, Reda Louis J

Аннотация

RATIONALE: This paper highlights the simplicity of interfacing an Atmospheric Solid Analysis Probe (ASAP) to a Linear Ion Trap Mass Spectrometer and shows that this technique can be used for the rapid generation of high-quality data from a range of sample types with minimal or no sample preparation.

METHODS: For a solid sample or surface deposit, the process entails rubbing a capillary melting tube a few times on the sample to transfer material to the capillary surface and then introducing it into the source of the mass spectrometer. Similarly, for a liquid sample, a capillary tube is dipped into the sample to just coat the surface or a few microliters may be applied to the tip of a capillary before being analyzed by Atmospheric Pressure Chemical Ionization in both positive and negative mode.

RESULTS: A rodenticide containing brodifacoum, black tar heroin and its impurities (morphine, codeine, noscapine, papaverine, and monoacetylmorphine), crack cocaine and 1-methylaminoanthraquinone dyestuff were successfully analyzed directly without any sample preparation. All compounds were detected using full scan mass spectrometry (MS), followed by confirmation by MS/MS. Preliminary results suggest that this technique could be used for quantitation.

CONCLUSIONS: Interfacing the ASAP to an ion trap mass spectrometer allows the ability to perform full scan, MS(n) experiments, and rapid positive/negative switching from a single sample introduction. Because of these features, this instrument is very useful for rapid, routine analysis and for confirmation with the use of in-house MS/MS libraries.