Rapid analysis of forensic samples using an atmospheric solid analysis probe interfaced to a linear ion trap mass spectrometer.
Аннотация
RATIONALE: This paper highlights the simplicity of interfacing an Atmospheric Solid Analysis Probe (ASAP) to a Linear Ion Trap Mass Spectrometer and shows that this technique can be used for the rapid generation of high-quality data from a range of sample types with minimal or no sample preparation.
METHODS
For a solid sample or surface deposit, the process entails rubbing a capillary melting tube a few times on the sample to transfer material to the capillary surface and then introducing it into the source of the mass spectrometer. Similarly, for a liquid sample, a capillary tube is dipped into the sample to just coat the surface or a few microliters may be applied to the tip of a capillary before being analyzed by Atmospheric Pressure Chemical Ionization in both positive and negative mode.
RESULTS
A rodenticide containing brodifacoum, black tar heroin and its impurities (morphine, codeine, noscapine, papaverine, and monoacetylmorphine), crack cocaine and 1-methylaminoanthraquinone dyestuff were successfully analyzed directly without any sample preparation. All compounds were detected using full scan mass spectrometry (MS), followed by confirmation by MS/MS. Preliminary results suggest that this technique could be used for quantitation.
CONCLUSIONS
Interfacing the ASAP to an ion trap mass spectrometer allows the ability to perform full scan, MS(n) experiments, and rapid positive/negative switching from a single sample introduction. Because of these features, this instrument is very useful for rapid, routine analysis and for confirmation with the use of in-house MS/MS libraries.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.