INTRODUCTION

Deep brain stimulation (DBS) is an effective treatment for medically refractory Parkinson's disease (PD). During DBS surgery, intraoperative testing is performed to confirm optimal lead placement by determining the stimulation thresholds for symptom improvement and side effects. However, the reliability of intraoperative testing in predicting distant postoperative thresholds is unknown. In this study, we hypothesized that intraoperative testing reliably estimates postoperative thresholds for both symptom improvement and side effects.

METHODS

We retrospectively analyzed a prospective database with intraoperative and postoperative thresholds for symptom improvement and side effects from a cohort of 66 PD patients who underwent STN DBS. We recorded the stimulation locations relative to the mid-commissural point. Within-patient stimulation pairs were generated by clustering the intraoperative stimulation locations closest to the DBS contacts. We computed the distance between stimulation locations and atlas-based pyramidal tract (PT) and medial lemniscus (ML) masks. A leave-one-out cross-validation analysis was performed to determine the reliability of intraoperative testing in predicting postoperative thresholds while controlling for the distance from the relevant tracks.

RESULTS

Intraoperative testing reliably predicted (area under ROC >0.8) postoperative thresholds for tremor and rigidity improvements, as well as stimulation-induced motor contractions and paresthesias. The reliability was poor for improvement in bradykinesia.

CONCLUSION

Intraoperative testing reliably predicts postoperative thresholds. These results are relevant during the informed consent process and patient counseling for DBS surgery. These will also guide the development of future methods for intraoperative feedback, especially during asleep DBS.

Для цитирования: Sammartino Francesco, Rege Rahul, Krishna Vibhor. Reliability of Intraoperative Testing During Deep Brain Stimulation Surgery.. Neuromodulation : journal of the International Neuromodulation Society. 2019. DOI: 10.1111/ner.13081.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.