A Novel External Quality Assessment of Chromosomal Karyotype Analysis Based on Chimera Image Assembly.

Weng Bing-Huan, Shen Xiao-Yun, Xu Ya-Li, Ying Jun, Xu Wei, Zou Duo-Bing, He Li-Mei, Jin Keqin, Li Lan-Juan

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

This study aimed to establish a new external quality assessment (EQA) of chromosomal karyotype analysis.

METHODS

Chimeric assembly A1 was established by collecting chimeric chromosome images prepared artificially from chromosomally abnormal amniocytes remaining after prenatal diagnosis. Chimeric assembly B1 and nonchimeric assembly C1 were constructed through the collection of chimeric and nonchimeric chromosome images from prenatal diagnosis, respectively. Then, chromosome images were selected randomly from assemblies A1, B1, or C1 to send to 20 technicians via email to verify the validity of a new EQA of chromosomal karyotype analysis.

RESULTS

According to the EQA of 20 technicians, 47,XX,+mar from assembly A was easily misdiagnosed as 47,XX,+19 or 47,XXY, and 45,XX,t(13;22) (q10;q10) was misdiagnosed as 45,XX,13S+,-22. The total misdiagnosis rate was 3.8%. For assembly B, 46,X,+mar and 46,X,idic(Y) were easily misdiagnosed as 46,XY and 46,X,+mar, respectively. In addition, some testers missed 47,XXX in 47,XXX[2]/46,XX[48], as well as 47,XX,+18 in 46,XX [47]/47,XX,+18[3], and 45,X and 47,XXX in 46,XX[47]/45,X[2]/47,XXX[1]. The total misdiagnosis rate was 4.2%. All karyo-types from assembly C were correctly diagnosed, although incorrect descriptions used for 4% of cases.

CONCLUSION

The quality of chromosome karyotype analysis can be comprehensively evaluated by a new EQA based on assembly A1 or B1.

Для цитирования: Weng Bing-Huan, Shen Xiao-Yun, Xu Ya-Li, Ying Jun, Xu Wei, Zou Duo-Bing, He Li-Mei, Jin Keqin, Li Lan-Juan. A Novel External Quality Assessment of Chromosomal Karyotype Analysis Based on Chimera Image Assembly.. Annals of clinical and laboratory science. 2020.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.