Efficacy of Noninvasive Ventilation With Expiratory Washout in Stable COPD Patients.

Kung Stacey, Semprini Alex C, Kirton Louis W, Fogarin Jess R, Zoellner Sascha K, Beasley Richard, Eathorne Allie, Semprini Ruth A C

СтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

Background

A noninvasive ventilation (NIV) mask has been designed to deliver NIV with expiratory washout to improve efficacy of ventilation by optimizing clearance of expired gases from the anatomic dead space. This study compared the performance and comfort of a novel investigational mask with expiratory washout with a conventional mask during NIV therapy.

Methods

In this pilot crossover study, participants with severe stable COPD attended a single visit to receive bi-level NIV through 2 masks; the investigational mask with expiratory washout and a conventional mask. The order of mask use was randomly allocated, and each mask was used for 60 min with a 30-60-min washout in between. The primary outcome was transcutaneous carbon dioxide at 60 min. Other physiologic and NIV device variables were also assessed.

Results

The mean difference (95% CI) in the transcutaneous carbon dioxide between the investigational and conventional masks at 60 min, adjusted for baseline, was -0.74 mm Hg, 95% CI -2.81 to 1.33 mm Hg (P = .45). The investigational mask with expiratory washout elicited a lower tidal volume (-128.7 mL, 95% CI -190.0 to -67.3 mL; P < .001) and minute ventilation (-2.28 L/min,, 95% CI -3.12 to -1.43 L/min; P < .001), and a higher leak (7.96 L/min, 95% CI 4.39-11.54 L/min; P < .001) than the conventional mask. There were no differences in other physiologic responses or ratings of dyspnea or comfort.

Conclusions

NIV therapy delivered by using a novel mask with expiratory washout was similarly effective at reducing transcutaneous carbon dioxide, whereas the delivered tidal volume and minute ventilation were significantly lower when compared with a conventional mask in participants with severe COPD.

Для цитирования: Kung Stacey, Semprini Alex C, Kirton Louis W, Fogarin Jess R, Zoellner Sascha K, Beasley Richard, Eathorne Allie, Semprini Ruth A C. Efficacy of Noninvasive Ventilation With Expiratory Washout in Stable COPD Patients.. Respiratory care. 2025. DOI: 10.1089/respcare.11876.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.