DOI 10.1097/ftd.0000000000001175
Analytical Validation of a Volumetric Absorptive Microsampling Method for Therapeutic Drug Monitoring of the Oral Targeted Anticancer Agents, Abiraterone, Alectinib, Cabozantinib, Imatinib, Olaparib, and Sunitinib, and Metabolites.
Аннотация
BACKGROUND
Volumetric Absorptive Microsampling (VAMS) is a useful tool for therapeutic drug monitoring (TDM) of oral targeted anticancer agents. VAMS aims to improve safety and efficacy by enabling at-home blood sample collection by patients. This study aimed to develop and validate an ultra-high performance liquid chromatography-tandem mass spectrometry method for the quantitative determination of abiraterone, alectinib, cabozantinib, imatinib, olaparib, sunitinib, and the metabolites, Δ(4)-abiraterone (D4A), alectinib-M4, imatinib-M1, and N -desethyl sunitinib, in dried whole blood samples using VAMS to support TDM.
METHODS
After the collection of 10 μL of whole blood sample using the VAMS device, the analytes were extracted from the tip using methanol with shaking, evaporated, and reconstituted in acetonitrile:0.1 mol/L ammonium hydroxide in water (1:1, vol/vol). The extracts were then analyzed using ultra-high performance liquid chromatography-tandem mass spectrometry. Validation experiments based on the ICH M10 guideline were carried out, and stability was evaluated under shipping and storage conditions. VAMS specimens were collected in the outpatient clinic to demonstrate the applicability of the assay.
RESULTS
The validated range of the method was considered accurate and precise for all analytes. Accordingly, the validation experiments met the relevant requirements, except for cross-analyte interference. Based on the stability data, shipment can be performed at room temperature within 14 days after sample collection and the VAMS specimen can be stored up to 9 months at -20 and -70°C. Samples from 59 patients were collected at the hospital.
CONCLUSIONS
The developed method could be used to successfully quantify the concentrations of abiraterone, D4A, alectinib, alectinib-M4, cabozantinib, imatinib, imatinib-M1, olaparib, sunitinib, and N -desethyl sunitinib within the validated range using VAMS. Therefore, the method can be used to estimate the dried whole blood-to-plasma ratios for TDM in the clinic.